Skip to content

ABL90 Flex

by Radiometer Medical ApS 5 variants

510(k) K120197

ABL90 Flex is a medical device (FDA Class II; FDA product code CHL, Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph) made by Radiometer Medical ApS, cleared under FDA 510(k) K120197 on 19 Mar 2012, active as of 9 Oct 2026. One FDA recall is linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 9 Oct 2026.

Regulatory status

as of registry sync 9 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK120197 Date19 Mar 2012ClearedStatusActive

Official records

From the FDA record

Received by the FDA
23 Jan 2012
FDA decision
19 Mar 2012 (Substantially Equivalent)
FDA review time
56 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K120197 (openFDA), as read on 9 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

3 recalls · FDA

    • Terminated Z-0010-2019 FDA Class II initiated 25 May 2018 on shared 510(k) K120197

    The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: The analyzer is in ready mode Operator lifts the inlet and then closes it again, thereby initiating a rinse Immediately after lifts the inlet and quickly closes it again

    • Terminated Z-1046-2015 FDA Class II initiated 8 Dec 2014 on shared 510(k) K120197

    The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.

    • Terminated Z-2164-2012 FDA Class II initiated 5 Jul 2012

    A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

4 dated events

  1. Cleared FDA 510(k) K120197
  2. Recall initiated FDA recalls Z-2164-2012
  3. Recall initiated FDA recalls Z-1046-2015
  4. Recall initiated FDA recalls Z-0010-2019

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
CHL
Device class
Class II
Regulation
21 CFR Part 862.1120
Advisory committee
Clinical Chemistry
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph).

Other devices from Radiometer Medical ApS

See all devices from Radiometer Medical ApS →