Equinoxe 314-01-15
Equinoxe 314-01-15 is a Class II device made by Exactech, Inc., cleared under FDA 510(k) K093430 on 2 Sep 2010 and listed in FDA GUDID since 30 Oct 2015 (UDI-DI 10885862084460), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10885862084460 | Date30 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK093430 | Date2 Sep 2010 | StatusActive |
Official records
- AccessGUDID record 10885862084460
- FDA 510(k) database K093430
Across registries
- FDA 510(k) K093430 decided 2 Sep 2010 Active
- FDA GUDID UDI-DI 10885862084460 listed 30 Oct 2015 Listed
Recalls
-
- Ongoing Z-1405-2024
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
Timeline
- Cleared FDA 510(k) K093430
- Listed FDA GUDID 10885862084460
- Recall initiated FDA recalls Z-1405-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Equinoxe
- Generic name
- Shoulder glenoid fossa prosthesis, prefabricated
- Model / Ref / Catalog
- 314-01-15
- Primary DI / UDI-DI
- 10885862084460
- 510(k) Number
- FDA product code
- KWT, KWS
- GMDN term
- Shoulder glenoid fossa prosthesis, prefabricated
- Device class
- Class II
- Regulation number
- 888.3650, 888.3660
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Shoulder glenoid fossa prosthesis, prefabricated.
- SMR TT Hybrid Glenoid 1379.59.111 by Limacorporate
- Humelock II Cemented Shoulder System 105-0002 by FX SOLUTIONS
- Humelock II Cemented Shoulder System 105-0001 by FX SOLUTIONS
- SMR TT Hybrid Glenoid 1379.59.210 by Limacorporate
- LINK Embrace Shoulder System 645-003/64 by WALDEMAR LINK GmbH & Co. KG
- Anatomical Shoulder™ 01.04214.340 by Zimmer, Inc.
- LINK Embrace Shoulder System 645-021/52 by WALDEMAR LINK GmbH & Co. KG
- Anatomical Shoulder™ 01.04214.400 by Zimmer, Inc.
- Comprehensive® Reverse Shoulder 406202 by Biomet Orthopedics, Inc.