GE Precision Mpi
by Nrt - Nordisk Røntgen Teknik A/S
510(k) K033486
GE Precision Mpi is a medical device (FDA Class II; FDA product code OWB, Interventional Fluoroscopic X-Ray System) made by Nrt - Nordisk Røntgen Teknik A/S, cleared under FDA 510(k) K033486 on 19 Nov 2003, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK033486 | Date19 Nov 2003 | StatusActive |
Official records
- FDA 510(k) database K033486
From the FDA record
- Received by the FDA
- 4 Nov 2003
- FDA decision
- 19 Nov 2003 (Substantially Equivalent)
- FDA review time
- 15 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-2123-2015
A reported incident of a monitor that fell from the suspension. A fall of a monitor could result in an injury to a patient or operator. Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.
-
- Terminated Z-0911-2014
GE Healthcare has recently become aware of a potential safety issue involving the Precision 500D, Precision MPI & Precision RXI X-ray imaging systems overhead LCD Video Monitor Suspension (Single monitor & Dual monitor). There is a potential for the locking nut to loosen out of the LCD Video Monitor Suspension yoke. This will not lead to an immediate fall of the suspended monitor since it will
-
- Terminated Z-1658-2009
The GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes of fluoroscopic irradiation time during an examination or procedure.
-
- Terminated Z-0753-2016
It was discovered that the Remote Touch Panel (RTP) of the GE Precision MPi X-ray system may not always boot up as intended and needs to be updated to properly accomplish its intended purpose.
-
- Terminated Z-2143-2008
- Terminated Z-2144-2008
- Terminated Z-2142-2008
The Precision MPi collimator did not contain a label identifying the device as being certified to comply with applicable requirements of the x-ray performance standard.
Timeline
- Cleared FDA 510(k) K033486
- Recall initiated FDA recalls Z-2143-2008
- Recall initiated FDA recalls Z-2144-2008
- Recall initiated FDA recalls Z-2142-2008
- Recall initiated FDA recalls Z-0753-2016
- Recall initiated FDA recalls Z-1658-2009
- Recall initiated FDA recalls Z-0911-2014
- Recall initiated FDA recalls Z-2123-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- OWB
- FDA applicant
- Nrt-Nordisk Rontgen Teknik A/S
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code OWB (Interventional Fluoroscopic X-Ray System).
- IntraSight Plus 797417 by Philips Image Guided Therapy Corporation
- PHILIPS MULTIDIAGNOST 4 by Philips Medical Systems
- ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono (VE40A) (ceiling); ARTIS pheno (VE40A) by SIEMENS MEDICAL SOLUTIONS USA, INC.
- MC2 Portable X-ray System by Oxos Medical
- Vascular Navigation PAD 2.0; Navigation Software Vascular PAD by Brainlab AG
- INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ by GE Healthcare
- LESSRAY by Saferay Spine, LLC
- CS-SERIES FLUOROSCOPY SYSTEM, CS-10 THROUGH CS-65 (11 CONFIGURATIONS) by Omega Medical Imaging, LLC
- INFISTITCH FOR I5 DIGITAL X-RAY IMAGING SYSTEM, MODEL I5 by INFIMED, INC.
- PHILIPS OMNIDIAGNOST by Philips Medical Systems