Intelli-C
510(k) K173631
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK173631 | Date11 Jul 2018 | StatusActive |
Recalls
-
- Ongoing Z-1784-2024
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
-
- Ongoing Z-1669-2023
- Ongoing Z-1668-2023
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
-
- Ongoing Z-0526-2022
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 26 Sep 2026
Intelli-C is an Interventional Fluoroscopic X-Ray System. It is a multipurpose tilt-C X-ray system designed for diagnostic and interventional procedures.
The device is supplied as a reusable system for use in radiology settings. It received FDA 510(k) clearance (K173631) as substantially equivalent to a predicate device.