Giraffe and Panda Bag and Mask Resuscitation System

by Ohmeda Medical

510(k) K070247

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK070247 Date20 Apr 2007ClearedStatusActive

Recalls

5 recalls · 1 ongoing · FDA

    • Ongoing Z-2357-2026 FDA Class I initiated 8 May 2026

    Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

  • 2 recalls, same reason

    • Terminated Z-0915-2014 FDA Class II initiated 12 Dec 2013
    • Terminated Z-0914-2014 FDA Class II initiated 12 Dec 2013

    Medical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.

  • 2 recalls, same reason

    • Terminated Z-1071-2014 FDA Class I initiated 1 Oct 2013
    • Terminated Z-0749-2013 FDA Class I initiated 15 Oct 2012

    Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
BTM
Device Class
Class II
Regulation
21 CFR Part 868.5915
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent