Giraffe and Panda Warmers
510(k) K123309
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK123309 | Date5 Feb 2013 | StatusActive |
Recalls
-
- Ongoing Z-2359-2026
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
-
- Terminated Z-1847-2019
- Terminated Z-1846-2019
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
-
- Terminated Z-0481-2019
- Terminated Z-0482-2019
- Terminated Z-0483-2019
- Terminated Z-0484-2019
Potential electric leakage affecting the user.
-
- Terminated Z-1550-2014
- Terminated Z-1547-2014
- Terminated Z-1548-2014
- Terminated Z-1549-2014
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
-
- Terminated Z-0912-2014
- Terminated Z-0913-2014
Medical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FMT
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5130
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent