Skip to content

IMACTIS CT-Navigation system

by Imactis SAS

510(k) K162314

IMACTIS CT-Navigation system is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Imactis SAS, cleared under FDA 510(k) K162314 on 24 Apr 2018, active as of 22 Oct 2026. 4 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK162314 Date24 Apr 2018ClearedStatusActive

Official records

From the FDA record

Received by the FDA
18 Aug 2016
FDA decision
24 Apr 2018 (Substantially Equivalent)
FDA review time
614 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K162314 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · 3 ongoing · FDA

    • Ongoing Z-1590-2026 FDA Class II initiated 3 Feb 2026

    There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.

    • Ongoing Z-2972-2020 FDA Class II initiated 24 Aug 2020

    The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.

    • Ongoing Z-0838-2020 FDA Class II initiated 1 Aug 2019

    The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

    • Terminated Z-1264-2021 FDA Class II initiated 11 Feb 2021

    Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory

Timeline

5 dated events

  1. Cleared FDA 510(k) K162314
  2. Recall initiated FDA recalls Z-0838-2020
  3. Recall initiated FDA recalls Z-2972-2020
  4. Recall initiated FDA recalls Z-1264-2021
  5. Recall initiated FDA recalls Z-1590-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAK
FDA applicant
Imactis, Sas
Device class
Class II
Regulation
21 CFR Part 892.1750
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K162314

6 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K162314
UDI-DISize and differences
IMACTIS® A10010
IMACTIS
A10010B681A100101
Prescription: yes; Single use: yes
On the market
IMACTIS® A10200
IMACTIS
A10200B681A102001
Prescription: yes; Single use: no
On the market
IMACTIS® I10100
IMACTIS
MA-I10100 / I10100B681I101002
Prescription: yes; Single use: yes
On the market
IMACTIS® J00099_US
IMACTIS
MA-J00099_US / J00099_USB681J00099US0
Prescription: no; Single use: no
On the market
IMACTIS® J00197
IMACTIS
J00197B681J001971
Prescription: no; Single use: no
On the market
IMACTIS® J02000
IMACTIS
MA-J02000 / J02000B681J020000
Prescription: yes; Single use: no
On the market

Similar devices

Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).

Other devices from Imactis SAS