IMACTIS CT-Navigation system
by Imactis SAS
510(k) K162314
IMACTIS CT-Navigation system is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Imactis SAS, cleared under FDA 510(k) K162314 on 24 Apr 2018, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK162314 | Date24 Apr 2018 | StatusActive |
Official records
- FDA 510(k) database K162314
From the FDA record
- Received by the FDA
- 18 Aug 2016
- FDA decision
- 24 Apr 2018 (Substantially Equivalent)
- FDA review time
- 614 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1590-2026
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
-
- Ongoing Z-2972-2020
The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.
-
- Ongoing Z-0838-2020
The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.
-
- Terminated Z-1264-2021
Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
Timeline
- Cleared FDA 510(k) K162314
- Recall initiated FDA recalls Z-0838-2020
- Recall initiated FDA recalls Z-2972-2020
- Recall initiated FDA recalls Z-1264-2021
- Recall initiated FDA recalls Z-1590-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- FDA applicant
- Imactis, Sas
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K162314
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| IMACTIS® A10010 | A10010 | B681A100101 | On the market | |
| IMACTIS® A10200 | A10200 | B681A102001 | On the market | |
| IMACTIS® I10100 | MA-I10100 / I10100 | B681I101002 | On the market | |
| IMACTIS® J00099_US | MA-J00099_US / J00099_US | B681J00099US0 | On the market | |
| IMACTIS® J00197 | J00197 | B681J001971 | On the market | |
| IMACTIS® J02000 | MA-J02000 / J02000 | B681J020000 | On the market |
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- SHIMADZU WHOLE BODY X-RAY COMPUTED TOMOGRAPHY SCANNER SCT-7000TS by Shimadzu Medical Systems
- PHILIPS TOMOSCAN SR 4000 by Philips Medical Systems North America, Inc.
- GE LIGHTSPEED ACT FP16 by GE MEDICAL SYSTEMS, INC.
- CSBP-001A BODY PERFUSION SYSTEM by Toshiba Medical Systems
- DENTASCAN by Mpdi, Inc.
- CT-GUIDE NEEDLE GUIDANCE SYSTEM by Activiews, Ltd.
- INSPACE 4D SOFTWARE PACKAGE WITH EXTENDED FUNCTIONALITY by Siemens Medi Cal Solutions, Inc.
- PHILIPS TOMOSCAN SR by Philips Medical Systems North America, Inc.
- XPEED COMPUTED TOMOGRAPHY X-RAY SYSTEM by Toshiba America Medical Systems, In.C
- EXEL 1002 CT SCANNER by Elscint, Inc.