Skip to content

Infant Warmer System

by Ohmeda Medical

510(k) K963058

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK963058 Date1 Oct 1996ClearedStatusActive

Recalls

3 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-2323-2017 FDA Class II initiated 5 May 2017
    • Terminated Z-1015-2017 FDA Class II initiated 4 Aug 2016

    GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

    • Terminated Z-1511-2017 FDA Class II initiated 18 Nov 2016

    Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FMT
Device Class
Class II
Regulation
21 CFR Part 880.5130
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Ohmeda Medical