Newest first, rows 1-2 of 2

  • Intelli-C Ref K173631 Open recall

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Adora DRFi (04550010) Ref K242948

    Active Regimes: 510(k) Registry: 510(k) Class:II

Company details

FDA FEI number
1000188474
Address
BIRKEGAARDSVEJ 16, HASSELAGER Midtjylland

FDA recalls4

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

FDA recall records list 4 recalls initiated by Nrt X-Ray A/S, the earliest on 2022-01-21 and the most recent on 2024-04-17. By recorded status: 4 ongoing. The reasons filed fall into these groups: software or firmware and mechanical integrity. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.

Generated automatically from FDA recall records, as a summary of the records listed below. It describes what the records state and is not an assessment of this manufacturer.

Recalls initiated per year

  • 2017 0
  • 2018 0
  • 2019 0
  • 2020 0
  • 2021 0
  • 2022 1
  • 2023 2
  • 2024 1
  • 2025 0
  • 2026 0
  • 2024

    • 17 Apr 2024 Z-1784-2024 Class II Ongoing X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator. Distribution: US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI. Submission number: K173631
  • 2023

    • 27 Apr 2023 Z-1668-2023 Class II Ongoing There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop. Distribution: US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY. Submission number: K173631
    • 27 Apr 2023 Z-1669-2023 Class II Ongoing There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop. Distribution: US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY. Submission number: K173631
  • 2022

    • 21 Jan 2022 Z-0526-2022 Class II Ongoing Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left. Distribution: USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden Submission number: K173631

Frequently asked questions

How many devices does Nrt X-Ray A/S have in the registry?

The registry currently lists 2 devices for Nrt X-Ray A/S. Devices recorded under merged duplicate company entries are counted once.

Which registries list Nrt X-Ray A/S's devices?

Among its 2 most recently registered devices, Nrt X-Ray A/S has devices listed in FDA 510(k) (2 devices).

What risk classes do Nrt X-Ray A/S's devices fall under?

Among its 2 most recently registered devices, the classes recorded for Nrt X-Ray A/S's devices are Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for Nrt X-Ray A/S?

FDA recall records list 4 recalls initiated by Nrt X-Ray A/S, the most recent initiated on 17 Apr 2024. 4 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from Nrt X-Ray A/S in the registry?

The most recently registered devices listed for Nrt X-Ray A/S include Intelli-C and Adora DRFi (04550010).

Generated automatically from registry data and not verified by a human reviewer.