Interstim® 357625
Interstim® 357625 is a Class III device made by Medtronic, Inc., approved under FDA PMA P080025 on 14 Mar 2011 (1 earlier submission) and listed in FDA GUDID since 23 Sep 2014 (UDI-DI 00643169036031), active as of 22 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00643169036031 | Date23 Sep 2014 | StatusListed |
| United States | FDA PMAAPPR | NumberP080025 | Date14 Mar 2011 | StatusActive |
Official records
- AccessGUDID record 00643169036031
- FDA PMA database P080025
- FDA PMA database P970004
Across registries
- FDA PMA P080025 decided 14 Mar 2011 Active
- FDA GUDID UDI-DI 00643169036031 listed 23 Sep 2014 Listed
Recalls
-
- Ongoing Z-1704-2025
Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.
-
- Ongoing Z-0372-2025
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
-
- Ongoing Z-2365-2023
Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.
-
- Terminated Z-0998-2022
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
-
- Terminated Z-1147-2015
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
-
- Terminated Z-0037-2008
A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .
-
- Terminated Z-0040-2008
Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).
Timeline
- Approved FDA PMA P970004
- Recall initiated FDA recalls Z-0037-2008
- Recall initiated FDA recalls Z-0040-2008
- Approved FDA PMA P080025
- Listed FDA GUDID 00643169036031
- Recall initiated FDA recalls Z-1147-2015
- Recall initiated FDA recalls Z-0998-2022
- Recall initiated FDA recalls Z-2365-2023
- Recall initiated FDA recalls Z-0372-2025
- Recall initiated FDA recalls Z-1704-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Interstim®
- Generic name
- Electrical-only medical device connection cable, reusable
- Model / reference
- 357625
- Primary DI / UDI-DI
- 00643169036031
- FDA product code
- EZW, QON
- GMDN term
- Electrical-only medical device connection cable, reusable
- Device class
- Class III
- Regulation number
- 876.5270
- GUDID distribution status
- In Commercial Distribution
- Size
- Length — 25.0 Centimeter
Similar devices in this category
Other devices filed under Electrical-only medical device connection cable, reusable.
- Guenter Bissinger Medizintechnik GmbH 80100042 by Günter Bissinger Medizintechnik GmbH
- 235mm Single Action Grasper 101-47034 by Surgical Innovations Ltd.
- NA DAC-NT by ST. JUDE MEDICAL, INC.
- Guenter Bissinger Medizintechnik GmbH 80100197 by Günter Bissinger Medizintechnik GmbH
- TRAUS ENDO Series ACL(B)-46EP by SAESHIN PRECISION CO.,LTD.
- TRAUS ENDO Series BAE03 by SAESHIN PRECISION CO.,LTD.
- NIBP Tube BPT3 by Edan Instruments,Inc.
- 305mm Single Action Grasper 101-48034 by Surgical Innovations Ltd.
- Bipolar cable SBCD-EU/1 by Soniquence, LLC
- ENT NIM Vital NIM4CPB2 by Medtronic Xomed Inc