Ivenix Infusion System 030-0027-06
Ivenix Infusion System 030-0027-06 (IMS) is a medical device (FDA Class II; FDA product code PHC, Infusion Safety Management Software) made by Fresenius Kabi, cleared under FDA 510(k) K183311 on 7 Jun 2019 and listed in FDA GUDID since 7 Aug 2023 (UDI-DI 00811505030443), active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
21 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00811505030443 | Date7 Aug 2023 | StatusListed |
| United States | FDA 510(k)SESE | NumberK183311 | Date7 Jun 2019 | StatusActive |
Official records
- AccessGUDID record 00811505030443
- FDA 510(k) database K183311
From the FDA record
- Device description
- Infusion Management System (IMS)
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K183311 decided 7 Jun 2019 Active
- FDA GUDID UDI-DI 00811505030443 listed 7 Aug 2023 Listed
Recalls
-
- Ongoing Z-2223-2026
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
-
- Ongoing Z-2224-2026
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
-
- Ongoing Z-2274-2026
Potential for the Cassette Loading Lever to break.
-
- Ongoing Z-1048-2026
Emphasizing instructions for LVP duration programming located in the IFU.
-
- Ongoing Z-0885-2026
Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
-
- Ongoing Z-0586-2026
Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
-
- Ongoing Z-0848-2026
Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
-
- Ongoing Z-0378-2026
Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.
-
- Ongoing Z-2232-2025
- Ongoing Z-2233-2025
- Ongoing Z-2234-2025
- Ongoing Z-2235-2025
Potential for external cassette leaks
-
- Ongoing Z-1869-2025
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
-
- Ongoing Z-1019-2025
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
-
- Ongoing Z-0763-2025
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
-
- Ongoing Z-3153-2024
The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
-
- Ongoing Z-2721-2024
A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
-
- Ongoing Z-1484-2024
Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
-
- Ongoing Z-0549-2024
The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.
-
- Ongoing Z-1313-2023
Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.
-
- Ongoing Z-2635-2023
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
-
- Completed Z-1283-2024
Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.
-
- Completed Z-1282-2024
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
-
- Terminated Z-0002-2023
The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.
-
- Terminated Z-1381-2022
Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.
Timeline
- Recall initiated FDA recalls Z-2234-2025
- Recall initiated FDA recalls Z-2235-2025
- Recall initiated FDA recalls Z-0378-2026
- Recall initiated FDA recalls Z-0586-2026
- Recall initiated FDA recalls Z-0848-2026
- Recall initiated FDA recalls Z-0885-2026
- Recall initiated FDA recalls Z-1048-2026
- Recall initiated FDA recalls Z-2274-2026
- Recall initiated FDA recalls Z-2223-2026
- Recall initiated FDA recalls Z-2224-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Ivenix Infusion System
- Generic name
- Infusion pump medication-dosage application software
- Model / Ref / Catalog
- 030-0027-06, IMS
- Primary DI / UDI-DI
- 00811505030443
- 510(k) Number
- FDA product code
- PHC
- GMDN term
- Infusion pump medication-dosage application software
- Device class
- Class II
- Regulation number
- 880.5725
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code PHC (Infusion Safety Management Software).
- BECTON DICKINSON INTELLIPORT SYSTEM by BECTON, DICKINSON & CO.
- LifeShield Infusion Safety Software Suite by ICU MEDICAL, INC.
- Vigilant Software Suite – Vigilant Master Med by Fresenius Kabi AG
- BD Intelliport System by Becton, Dickinson and Company
- BD Intelliport System; BD Intelliport Reader (516230); BD Intelliport Sensor (516229); BD Intelliport Gateway (516232); BD Intelliport Mount (516233); BD Intelliport Charger (516231) by BECTON, DICKINSON & CO.
- MEDFUSION 4000 PUMP DEVICE SPECIFIC REPORTS SOLFWARE FOR PHARMGUARD SERVER VERSION 2.0 by Smiths Medical Asd, Inc.