Laguna Pedicle Screw System
510(k) K050060
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK050060 | Date4 May 2005 | StatusActive |
Recalls
-
- Terminated Z-0153-2009
This action is being taken voluntarily as a result of a Warning Letter issued to AlIez Spine LLC by the Food & Drug Administration. The Company decided it could better address the FDA's concerns regarding Current Good Manufacturing Practice requirements of the Quality System without any product in the market so it could focus all of its efforts on responding to the FDA.
-
- Terminated Z-1819-2008
This field action was initiated because the cleaning, decontamination and sterilization procedures specified in the "Instructions for Use" for the impacted systems are incorrect. The use of these parameters in performing cleaning, decontamination and sterilization procedures may not result in clean and sterile product.
-
- Terminated Z-2065-2008
- Terminated Z-2069-2008
- Terminated Z-2063-2008
- Terminated Z-2076-2008
- Terminated Z-2067-2008
- Terminated Z-2061-2008
- Terminated Z-2062-2008
- Terminated Z-2075-2008
- Terminated Z-2077-2008
- Terminated Z-2060-2008
- Terminated Z-2072-2008
- Terminated Z-2074-2008
- Terminated Z-2064-2008
- Terminated Z-2071-2008
- Terminated Z-2078-2008
- Terminated Z-2070-2008
- Terminated Z-2066-2008
- Terminated Z-2073-2008
- Terminated Z-2068-2008
This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
-
- Terminated Z-0189-2008
- Terminated Z-0190-2008
- Terminated Z-0191-2008
- Terminated Z-0187-2008
- Terminated Z-0188-2008
- Terminated Z-0186-2008
Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MNI
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3070
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent