Allez Spine, LLC
United StatesFDA applicant, unverified
- Devices
- 7
- FDA recalls
- 30
At a glance
Market reach
- United States
- FDA recalls
What Allez Spine, LLC makes
Allez Spine, LLC manufactures spinal fixation hardware, including monoaxial and polyaxial pedicle screw systems, cross connectors, and size-specific pedicle screws designed for spinal stabilization procedures. The Laguna and Del Mar series are intended for use in posterior spinal fusion surgeries, providing rigid or flexible fixation options depending on the screw type.
The company’s devices are classified as Class II under FDA regulations and are supported by 510(k) clearances. All products are listed in the FDA’s 510(k) database. The manufacturer is based in the United States.
Catalogue
Laguna Monoaxial Pedicle Screw System Ref K051275
Regimes: 510(k) Registry: 510(k)Allez Spine Cross Connectors Ref K070460
Regimes: 510(k) Registry: 510(k)Laguna Size 8 Pedicle Screw Ref K072678
Regimes: 510(k) Registry: 510(k)Laguna Polyaxial Pedicle Screw System Ref K083826
Regimes: 510(k) Registry: 510(k)Laguna Size 8 Pedicle Screw System Ref K091995
Regimes: 510(k) Registry: 510(k)Del Mar Monoaxial Pedicle Screw System Ref K091219
Regimes: 510(k) Registry: 510(k)Laguna Pedicle Screw System Ref K050060
Regimes: 510(k) Registry: 510(k)
FDA recalls
2008
- 21 Aug 2008 Z-0153-2009 Unclassified (pre-2012) Terminated This action is being taken voluntarily as a result of a Warning Letter issued to AlIez Spine LLC by the Food & Drug Administration. The Company decided it could better address the FDA's concerns regarding Current Good Manufacturing Practice requirements of the Quality System without any product in the market so it could focus all of its efforts on responding to the FDA.
- 21 Aug 2008 Z-0154-2009 Unclassified (pre-2012) Terminated This action is being taken voluntarily as a result of a Warning Letter issued to AlIez Spine LLC by the Food & Drug Administration. The Company decided it could better address the FDA's concerns regarding Current Good Manufacturing Practice requirements of the Quality System without any product in the market so it could focus all of its efforts on responding to the FDA.
- 21 Aug 2008 Z-0155-2009 Unclassified (pre-2012) Terminated This action is being taken voluntarily as a result of a Warning Letter issued to AlIez Spine LLC by the Food & Drug Administration. The Company decided it could better address the FDA's concerns regarding Current Good Manufacturing Practice requirements of the Quality System without any product in the market so it could focus all of its efforts on responding to the FDA.
- 14 Apr 2008 Z-1819-2008 Unclassified (pre-2012) Terminated This field action was initiated because the cleaning, decontamination and sterilization procedures specified in the "Instructions for Use" for the impacted systems are incorrect. The use of these parameters in performing cleaning, decontamination and sterilization procedures may not result in clean and sterile product.
- 14 Apr 2008 Z-1820-2008 Unclassified (pre-2012) Terminated This field action was initiated because the cleaning, decontamination and sterilization procedures specified in the "Instructions for Use" for the impacted systems are incorrect. The use of these parameters in performing cleaning, decontamination and sterilization procedures may not result in clean and sterile product.
2007
- 7 Nov 2007 Z-2060-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2061-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2062-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2063-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2064-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
Show the 25 most recent of 30 recalls
2007
- 7 Nov 2007 Z-2065-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2066-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2067-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2068-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2069-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2070-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2071-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2072-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2073-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2074-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2075-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2076-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2077-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 7 Nov 2007 Z-2078-2008 Unclassified (pre-2012) Terminated This recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
- 30 Jul 2007 Z-0186-2008 Unclassified (pre-2012) Terminated Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
Frequently asked questions
How many devices does Allez Spine, LLC have in the registry?
The registry currently lists 7 devices for Allez Spine, LLC. Devices recorded under merged duplicate company entries are counted once.
Which registries list Allez Spine, LLC's devices?
Among its 7 most recently registered devices, Allez Spine, LLC has devices listed in FDA 510(k) (7 devices).
What risk classes do Allez Spine, LLC's devices fall under?
Among its 7 most recently registered devices, the classes recorded for Allez Spine, LLC's devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Allez Spine, LLC?
FDA recall records list 30 recalls initiated by Allez Spine, LLC, the most recent initiated on 21 Aug 2008. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Allez Spine, LLC in the registry?
The most recently registered devices listed for Allez Spine, LLC include LAGUNA MONOAXIAL PEDICLE SCREW SYSTEM, ALLEZ SPINE CROSS CONNECTORS, LAGUNA SIZE 8 PEDICLE SCREW and LAGUNA POLYAXIAL PEDICLE SCREW SYSTEM.