Mahurkar 8888345629
Mahurkar 8888345629 is a triple-lumen haemodialysis catheter (FDA Class II; FDA product code NIE, Catheter, Hemodialysis, Triple Lumen, Non-Implanted) made by Mozarc Medical US LLC, cleared under FDA 510(k) K192302 on 17 Jan 2020 (1 earlier submission) and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 10884521006799), active as of 23 Oct 2026. 32 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
31 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SEKD | NumberK192302 | Date17 Jan 2020 | StatusActive |
| United States | GUDID listing | Number10884521006799 | Date24 Sep 2016 | StatusListed |
Official records
- AccessGUDID record 10884521006799
- FDA 510(k) database K192302
- FDA 510(k) database K020089
From the FDA record
- Device description
- Acute Triple Lumen Catheter Kit,Curved Extensions
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K192302 decided 17 Jan 2020 Active
- FDA GUDID UDI-DI 10884521006799 listed 24 Sep 2016 Listed
Recalls
-
- Ongoing Z-0421-2026
- Ongoing Z-0422-2026
- Ongoing Z-0423-2026
- Ongoing Z-0424-2026
- Ongoing Z-0425-2026
- Ongoing Z-0426-2026
A potential issue with the seal integrity of header bag packaging.
-
- Ongoing Z-2388-2023
- Ongoing Z-2367-2023
- Ongoing Z-2368-2023
- Ongoing Z-2369-2023
- Ongoing Z-2370-2023
- Ongoing Z-2371-2023
- Ongoing Z-2372-2023
- Ongoing Z-2373-2023
- Ongoing Z-2374-2023
- Ongoing Z-2375-2023
- Ongoing Z-2376-2023
- Ongoing Z-2377-2023
- Ongoing Z-2378-2023
- Ongoing Z-2379-2023
- Ongoing Z-2380-2023
- Ongoing Z-2381-2023
- Ongoing Z-2382-2023
- Ongoing Z-2383-2023
- Ongoing Z-2384-2023
- Ongoing Z-2385-2023
- Ongoing Z-2386-2023
- Ongoing Z-2387-2023
- Ongoing Z-2389-2023
- Ongoing Z-2390-2023
- Ongoing Z-2391-2023
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
-
- Terminated Z-1955-2019
The priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.
Timeline
- Recall initiated FDA recalls Z-2387-2023
- Recall initiated FDA recalls Z-2389-2023
- Recall initiated FDA recalls Z-2390-2023
- Recall initiated FDA recalls Z-2391-2023
- Recall initiated FDA recalls Z-0421-2026
- Recall initiated FDA recalls Z-0422-2026
- Recall initiated FDA recalls Z-0423-2026
- Recall initiated FDA recalls Z-0424-2026
- Recall initiated FDA recalls Z-0425-2026
- Recall initiated FDA recalls Z-0426-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Mahurkar
- Generic name
- Triple-lumen haemodialysis catheter
- Model / reference
- 8888345629
- Primary DI / UDI-DI
- 10884521006799
- FDA product code
- NIE
- GMDN term
- Triple-lumen haemodialysis catheter
- Device class
- Class II
- Regulation number
- 876.5540
- GUDID distribution status
- In Commercial Distribution
- Size
- Length — 20 CentimeterCatheter Gauge — 12 French