Mammotome MKEY1 is a mammographic stereotactic biopsy system (FDA Class II; FDA product code KNW, Instrument, Biopsy) made by Devicor Medical Products, Inc., cleared under FDA 510(k) K033700 on 20 Feb 2004 and listed in FDA GUDID since 13 Dec 2016 (UDI-DI 00841911100560), active as of 22 Oct 2026. One FDA recall is linked to it.
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00841911100560 | Date13 Dec 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK033700 | Date20 Feb 2004 | StatusActive |
Official records
- AccessGUDID record 00841911100560
- FDA 510(k) database K033700
From the FDA record
- Device description
- Remote Keypad
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K033700 decided 20 Feb 2004 Active
- FDA GUDID UDI-DI 00841911100560 listed 13 Dec 2016 Listed
Recalls
-
- Terminated Z-0440-05
A capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.
Timeline
- Cleared FDA 510(k) K033700
- Recall initiated FDA recalls Z-0440-05
- Listed FDA GUDID 00841911100560
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Mammotome
- Generic name
- Mammographic stereotactic biopsy system
- Model / Ref / Catalog
- MKEY1
- Primary DI / UDI-DI
- 00841911100560
- 510(k) Number
- FDA product code
- KNW
- GMDN term
- Mammographic stereotactic biopsy system
- Device class
- Class II
- Regulation number
- 876.1075
- GUDID distribution status
- In Commercial Distribution