Medfusion 4000-0106-00

by ICU Medical, Inc.

Medfusion 4000-0106-00 is a bedside syringe pump (FDA Class II) made by ICU Medical, Inc., cleared under FDA 510(k) K111386 on 29 Aug 2011 and listed in FDA GUDID since 7 Feb 2017 (UDI-DI 15019517070750), active as of 23 Oct 2026. 15 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

9 ongoing

This device is named in the recall

Documents

510(k) K111386

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number15019517070750 UDI-DI Date7 Feb 2017ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK111386 Date29 Aug 2011DecidedStatusActive

Official records

From the FDA record

MRI safety
MR Conditional
Natural rubber latex
Labelled as not made with natural rubber latex
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K111386 decided 29 Aug 2011 Active
  • FDA GUDID UDI-DI 15019517070750 listed 7 Feb 2017 Listed

Recalls

16 recalls · 9 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1541-2024 FDA Class II initiated 19 Dec 2023 UDI-DI 15019517070750 listed
    • Ongoing Z-1543-2024 FDA Class II initiated 19 Dec 2023

    Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

    • Ongoing Z-0864-2024 FDA Class I initiated 19 Dec 2023

    Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.

    • Ongoing Z-0079-2024 FDA Class II initiated 16 Aug 2023 UDI-DI 15019517070750 listed

    A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

    • Ongoing Z-0272-2023 FDA Class II initiated 10 Oct 2022

    Potential for delay in sending Infusion Start Messages to third-party Electronic Health Record systems which may affect infusion auto-programming on the pump and documentation

    • Ongoing Z-1271-2022 FDA Class I initiated 18 Apr 2022

    Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001

  • 2 recalls, same reason

    • Ongoing Z-2734-2020 FDA Class I initiated 26 Jun 2020
    • Ongoing Z-2735-2020 FDA Class I initiated 26 Jun 2020

    Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.

    • Ongoing Z-0653-2022 FDA Class II initiated 3 Dec 2015

    There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

    • Terminated Z-0610-2020 FDA Class I initiated 28 Oct 2019

    There is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.

    • Terminated Z-2554-2019 FDA Class II initiated 15 Oct 2018 on shared 510(k) K111386

    Customers who utilize PharmGuard Server Software (PGS) with Medfusion 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion 4000 pumps.

    • Terminated Z-2662-2018 FDA Class II initiated 13 Apr 2018

    Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.

    • Terminated Z-1134-2018 FDA Class II initiated 13 Nov 2017

    Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.

    • Terminated Z-0313-2018 FDA Class II initiated 22 Aug 2017

    Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.

    • Terminated Z-0540-2014 FDA Class II initiated 22 Nov 2013

    Smiths Medical is conducting a recall involving all Medfusion 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po

    • Terminated Z-2107-2012 FDA Class I initiated 14 Feb 2012

    Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

18 dated events

  1. Recall initiated FDA recalls Z-2554-2019
  2. Recall initiated FDA recalls Z-0610-2020
  3. Recall initiated FDA recalls Z-2734-2020
  4. Recall initiated FDA recalls Z-2735-2020
  5. Recall initiated FDA recalls Z-1271-2022
  6. Recall initiated FDA recalls Z-0272-2023
  7. Recall initiated FDA recalls Z-0079-2024
  8. Recall initiated FDA recalls Z-1541-2024
  9. Recall initiated FDA recalls Z-0864-2024
  10. Recall initiated FDA recalls Z-1543-2024

8 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Medfusion
Generic name
Bedside syringe pump
Model / reference
4000-0106-00
Primary DI / UDI-DI
15019517070750
510(k) Number
FDA product code
MRZ, FRN
GMDN term
Bedside syringe pump
Device class
Class II
Regulation number
880.5725
GUDID distribution status
In Commercial Distribution
MRI safety
MR Conditional

Similar devices

Other devices filed under FDA product code MRZ (ACCESSORIES, PUMP, INFUSION).

Other devices from ICU Medical, Inc.

See all devices from ICU Medical, Inc. →

Properties

  • Serial number
Other properties (14, all answered No)