microTargeting™ Drive Array Clamp 66-AC-NC-00465
by Fhc, Inc.
microTargeting Drive Array Clamp 66-AC-NC-00465 is a neurological stereotactic surgery system (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Fhc, Inc., cleared under FDA 510(k) K011992 on 14 Aug 2001 and listed in FDA GUDID since 17 Sep 2016 (UDI-DI 00873263004654), active as of 22 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00873263004654 | Date17 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK011992 | Date14 Aug 2001 | StatusActive |
Official records
- AccessGUDID record 00873263004654
- FDA 510(k) database K011992
From the FDA record
- Device description
- Non shilded array clamp for use with microTargeting Array Carrier
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use
Across registries
- FDA 510(k) K011992 decided 14 Aug 2001 Active
- FDA GUDID UDI-DI 00873263004654 listed 17 Sep 2016 Listed
Recalls
-
- Terminated Z-2353-2018
- Terminated Z-2359-2018
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
-
- Terminated Z-0486-2009
Measuring fixture is incorrectly graduated.
-
- Terminated Z-0179-06
- Terminated Z-0180-06
Power Assist microTargeting System may advance beyond intended target
-
- Terminated Z-0895-05
Potential for non-sterile pin to contaminate sterile field
Timeline
- Cleared FDA 510(k) K011992
- Recall initiated FDA recalls Z-0895-05
- Recall initiated FDA recalls Z-0179-06
- Recall initiated FDA recalls Z-0180-06
- Recall initiated FDA recalls Z-0486-2009
- Listed FDA GUDID 00873263004654
- Recall initiated FDA recalls Z-2353-2018
- Recall initiated FDA recalls Z-2359-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- microTargeting™ Drive Array Clamp
- Generic name
- Neurological stereotactic surgery system
- Model / Ref / Catalog
- 66-AC-NC-00465, 66-AC-NC
- Primary DI / UDI-DI
- 00873263004654
- 510(k) Number
- FDA product code
- HAW
- GMDN term
- Neurological stereotactic surgery system
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- Stealth Autoguide™ System 29439 by Medtronic Navigation Inc.
- SNT DISPOSABLE BIOPSY NEEDLE by Surgical Navigation Technologies, Inc.
- IPLAN! FLOW by Brainlab AG
- 6000-060-000 by Stryker Leibinger
- Autoguide 27563 by Medtronic Navigation Inc.
- KELLY-GROERSS-KALL CT STEROTACTIC ADAPTATION SYS by Mayo Clinic
- Cranial EM System by Brainlab AG
- ORTHOPILOT by Kinamed, Inc.
- GE HEALTHCARE INSTA TRAK 3500PLUS/ FLUORO TRAK 9800 PLUS SYSTEM WITH AXCESS BONE PIN, MODEL PART # 1006955 by GE HealthCare Technologies Inc.
- SNN SCOUT by I.S.G. Technologies, Inc.