Skip to content

microTargeting™ Insertion Tube 66-IT-IT2P-00769

by Fhc, Inc. GUDID

Listed FDA GUDID listing, no authorization on record Open recall

Regulatory status

as of 29 Oct 2026

Recall ongoing. 1 ongoing FDA recall names this device. See recalls

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00873263007693 UDI-DIDate8 May 2026ListedStatusListed GUDID: in commercial distribution

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Recalls

8 recalls · 1 ongoing · FDA

    • Ongoing Z-0312-2025 FDA Class I initiated 4 Oct 2024

    A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

  • 6 recalls, same reason

    • Terminated Z-2346-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2355-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2348-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2356-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2347-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2345-2018 FDA Class II initiated 10 May 2018

    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

    • Terminated Z-0895-05 FDA Unclassified (pre-2012) initiated 14 Mar 2005

    Potential for non-sterile pin to contaminate sterile field

Specification

GUDID

Trade Name
microTargeting™ Insertion Tube
Generic Name
Intracerebral cannula, nonimplantable, single-use
Model / Ref / Catalog
66-IT-IT2P-00769, 66-IT-IT2P
Primary DI / UDI-DI
00873263007693
510(k) Number
FDA Product Code
HAW
GMDN Term
Intracerebral cannula, nonimplantable, single-use
Device Class
Class II
Regulation Number
882.4560
GUDID distribution status
In Commercial Distribution
Sterilization Method
Ethylene Oxide

Other devices from Fhc, Inc.

See all devices from Fhc, Inc. →

Properties

  • Sterile
  • Single Use
  • Prescription Use
  • Lot / Batch Number
  • Expiration Date
  • Manufacturing Date

Other properties (8)

  • Implantable: No
  • Contains Latex: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No