Mini 6800 Digital Mobile C-Arm

by Ge Dec Medical Systems

510(k) K992506

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

7 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Abbreviated NumberK992506 Date19 Aug 1999ClearedStatusActive

Recalls

7 recalls · FDA

    • Terminated Z-0974-2017 FDA Class II initiated 19 Dec 2016

    GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.

    • Terminated Z-0193-2015 FDA Class II initiated 3 Oct 2014

    GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.

    • Terminated Z-0851-2007 FDA Unclassified (pre-2012) initiated 12 Jun 2007 on shared 510(k) K992506

    X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma.

    • Terminated Z-0928-2007 FDA Unclassified (pre-2012) initiated 24 May 2007 on shared 510(k) K992506

    X-ray systems did not have the required labeling to indicate Air Kerma Rate and cumulative Air Kerma during fluoroscopic procedures

    • Terminated Z-2126-2012 FDA Class II initiated 27 Mar 2007 on shared 510(k) K992506

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and

    • Terminated Z-2040-2012 FDA Class I initiated 20 Nov 2006

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo

    • Terminated Z-0704-06 FDA Unclassified (pre-2012) initiated 28 Mar 2006 on shared 510(k) K992506

    Radiation exposure rate could exceed specifications during fluoroscopic x-ray procedure.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
OXO
Device Class
Class II
Regulation
21 CFR Part 892.1650
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Ge Dec Medical Systems

See all devices from Ge Dec Medical Systems →