Series 9600 Mobile Digital Imaging System
510(k) K963952
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK963952 | Date23 Dec 1996 | StatusActive |
Recalls
-
- Terminated Z-1709-2008
Secondary collimator filters may be missing on certain x-ray units. It was discovered that the secondary collimator filter on some OEC 9600 C-arms was omitted when a collimator assembly was replaced. This would reduce the filtration and therefore the Half-Value-Layer (HVL) of the X-ray beam, and may result in non-compliance with respect to Maximum Entrance Exposure Rate (EER) unless the maximum EE
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent