Skip to content

Modification to: Apexpro Telemetry System

by GE Medical Systems Information Technologies

510(k) K032369

Modification to: Apexpro Telemetry System is a medical device (FDA Class II; FDA product code MHX, Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)) made by GE Medical Systems Information Technologies, cleared under FDA 510(k) K032369 on 15 Aug 2003, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

Recall on shared 510(k) K032369; other models may be affected

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK032369 Date15 Aug 2003ClearedStatusActive

Official records

From the FDA record

Received by the FDA
1 Aug 2003
FDA decision
15 Aug 2003 (Substantially Equivalent)
FDA review time
14 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K032369 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · 1 ongoing · FDA

    • Ongoing Z-0175-2020 FDA Class II initiated 30 Aug 2019 on shared 510(k) K032369

    May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.

    • Terminated Z-0725-2020 FDA Class II initiated 12 Nov 2019 on shared 510(k) K032369

    When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.

    • Terminated Z-1227-2009 FDA Unclassified (pre-2012) initiated 24 Feb 2009

    Intermittent or continuous loss of patient monitoring.

    • Terminated Z-0055-2007 FDA Unclassified (pre-2012) initiated 4 Aug 2006

    When a patient being monitored with ApexPro/Apex Pro CH Telemetry with ST monitoring disabled is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC (Clinical Information Center) does not occur.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

5 dated events

  1. Cleared FDA 510(k) K032369
  2. Recall initiated FDA recalls Z-0055-2007
  3. Recall initiated FDA recalls Z-1227-2009
  4. Recall initiated FDA recalls Z-0175-2020
  5. Recall initiated FDA recalls Z-0725-2020

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
MHX
Device class
Class II
Regulation
21 CFR Part 870.1025
Advisory committee
Cardiovascular
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code MHX (Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)).

Other devices from GE Medical Systems Information Technologies, Inc.

See all devices from GE Medical Systems Information Technologies, Inc. →