Modification to: Apexpro Telemetry System
by GE Medical Systems Information Technologies
510(k) K032369
Modification to: Apexpro Telemetry System is a medical device (FDA Class II; FDA product code MHX, Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)) made by GE Medical Systems Information Technologies, cleared under FDA 510(k) K032369 on 15 Aug 2003, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK032369 | Date15 Aug 2003 | StatusActive |
Official records
- FDA 510(k) database K032369
From the FDA record
- Received by the FDA
- 1 Aug 2003
- FDA decision
- 15 Aug 2003 (Substantially Equivalent)
- FDA review time
- 14 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0175-2020
May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.
-
- Terminated Z-0725-2020
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
-
- Terminated Z-1227-2009
Intermittent or continuous loss of patient monitoring.
-
- Terminated Z-0055-2007
When a patient being monitored with ApexPro/Apex Pro CH Telemetry with ST monitoring disabled is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC (Clinical Information Center) does not occur.
Timeline
- Cleared FDA 510(k) K032369
- Recall initiated FDA recalls Z-0055-2007
- Recall initiated FDA recalls Z-1227-2009
- Recall initiated FDA recalls Z-0175-2020
- Recall initiated FDA recalls Z-0725-2020
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MHX
- Device class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory committee
- Cardiovascular
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code MHX (Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)).
- DeltaXL Monitor MS18596 by Draeger Medical Systems, Inc.
- PARAMETERS MONITORES by Ivy Biomedical Systems, Inc.
- IntelliVue Patient Monitors MX100, MX400, MX430, MX450, MX500, MX550, IntelliVue Patient Monitors MX700, MX800, IntelliVue Patient Monitors MP5, MP5SC, Multi-Measurement Module X3 by Philips Medizin Systeme Boeblingen GmbH
- M540 Hemo Connect Cable, 2.9 m MS23898 by Draeger Medical Systems, Inc.
- MP5 INTELLIVUE PATIENT MONITOR by Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
- Patient Cable 9293-033-52 by MORTARA INSTRUMENT, INC.
- DATEX-OHMEDA S/5 COMPACT CRITICAL CARE MONITOR WITH L-CICU05 AND L-CICU05A SOFTWARE by GE Healthcare
- IntelliVue MP5 Patient Monitor by Philips Medizin Systeme Böblingen GmbH
- PATIENT MONITOR by Advanced Instruments, LLC
- PHILIPS MULTI-PARAMETER MODULE, MODEL M3001A; PHILIPS COMPONENT COMPACT MONITOR, RELEASE A.02, MODEL M1275B by Philips Medical Systems
Other devices from GE Medical Systems Information Technologies, Inc.
See all devices from GE Medical Systems Information Technologies, Inc. →