Modification to Hi-Art Treatment System
510(k) K082005
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK082005 | Date25 Aug 2008 | StatusActive |
Recalls
-
- Terminated Z-3108-2011
As a result of an internal review, TomoTherapy has identified an issue with HiArt versions 4.0.x, and HD versions 1.0.x that we would like to bring to your attention. During DICOM export of plan level images with a non square exported Field of View (FOV), an anomaly in the process of squaring the plan level image may cause the image to shift with respect to ROIs and dose. When the anomaly occu
-
- Terminated Z-3163-2011
TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the TomoMobile Hi-Art System. TomoTherapy has discovered that when attempting to open the TomoMobile shielding doors, while the door hinges are in the locked position, the hinges may fail allowing the door to disengage from the shielding.
-
- Terminated Z-1649-2011
It was determined that the Treatment Planning Station (TPS) can potentially under dose.
-
- Terminated Z-0814-2011
TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the Hi-Art System and make it so the user cannot generate a Completion Procedure.
-
- Terminated Z-2072-2010
An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
-
- Terminated Z-2115-2010
In some cases, the patient's diagnostic CT image is narrower than the Hi-Art radiotherapy couch image.
-
- Terminated Z-2219-2010
An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect. The Planning Station Plan Settings and DQA Setup tabs allow for images to be moved during planning. During registration when roll is applied on moved images, the Operator Station incorrectly rolls the image about
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent