Tomotherapy Treatment System Fast Optimizer Ref K112776
Regimes: 510(k) Registry: 510(k)Modification To: Hi-Art System Ref K060912
Regimes: 510(k) Registry: 510(k)Modification To: Hi-Art System Ref K033347
Regimes: 510(k) Registry: 510(k)Modification To Hi-Art Treatment System Ref K082005
Regimes: 510(k) Registry: 510(k)Hi-Art System Ref K042739
Regimes: 510(k) Registry: 510(k)Tomotherapy Treatment System Ref K112446
Regimes: 510(k) Registry: 510(k)
FDA recalls
2014
- 21 Nov 2014 Z-0507-2015 Class II Terminated Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.
Full distribution text
- 17 Jul 2014 Z-2239-2014 Class II Terminated Accuray is voluntarily recalling TomoTherapy H Series software versions 1.2.0, 1.2.1 and 1.2.2 (Hi-Art 4.2.0, 4.2.1 and 4.2.2).
Accuray has identified potential safety issues (anomalies) with these software versions.
Full distribution text
- 16 Jul 2014 Z-2240-2014 Class II Terminated Accuray is voluntarily recalling TomoTherapy H Series software versions 2.0.0 and 2.0.1 (Hi-Art 5.0.0 and 5.0.1). Accuray has identified potential safety issues (anomalies) with these software versions.
Full distribution text
- 21 Nov 2014 Z-0507-2015 Class II Terminated Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.
2011
- 26 Jan 2011 Z-1649-2011 Unclassified (pre-2012) Terminated It was determined that the Treatment Planning Station (TPS) can potentially under dose.
Full distribution text
- 26 Jan 2011 Z-1649-2011 Unclassified (pre-2012) Terminated It was determined that the Treatment Planning Station (TPS) can potentially under dose.
2010
- 8 Nov 2010 Z-0814-2011 Unclassified (pre-2012) Terminated TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the Hi-Art System and make it so the user cannot generate a Completion Procedure.
Full distribution text
- 23 Jun 2010 Z-2072-2010 Unclassified (pre-2012) Terminated An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer.
With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
Full distribution text
- 22 Feb 2010 Z-2115-2010 Unclassified (pre-2012) Terminated In some cases, the patient's diagnostic CT image is narrower than the Hi-Art radiotherapy couch image.
- 8 Nov 2010 Z-0814-2011 Unclassified (pre-2012) Terminated TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the Hi-Art System and make it so the user cannot generate a Completion Procedure.
2009
- 1 Dec 2009 Z-2219-2010 Unclassified (pre-2012) Terminated An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect. The Planning Station Plan Settings and DQA Setup tabs allow for images to be moved during planning. During registration when roll is applied on moved images, the Operator Station incorrectly rolls the image about
- 10 Aug 2009 Z-2243-2009 Unclassified (pre-2012) Terminated TomoTherapy received one customer report indicating that after starting a patient procedure from the Status
Console, the radiation on light and the audible indicator on the Status Console came on as expected,
however the Operator Station (OS) did not display the dose rate and the procedure time, or the couch position updates. Within a few minutes, the customer pressed the Stop button on the Sta
Full distribution text
2008
- 14 Oct 2008 Z-0628-2009 Unclassified (pre-2012) Terminated TomoTherapy Incorporated identified a potential issue with the Hi-Art system during the course of ongoing testing.
The Operator Station Calibration panel provides access to view and modify machine specific configuration settings. Access to these settings has always been restricted to individuals with appropriate security rights, being limited to only the "Superuser" and "Field Service engine
Full distribution text
- 14 Oct 2008 Z-0628-2009 Unclassified (pre-2012) Terminated TomoTherapy Incorporated identified a potential issue with the Hi-Art system during the course of ongoing testing.
The Operator Station Calibration panel provides access to view and modify machine specific configuration settings. Access to these settings has always been restricted to individuals with appropriate security rights, being limited to only the "Superuser" and "Field Service engine
Show all 12 recalls
2008
- 18 Sep 2008 Z-0339-2009 Unclassified (pre-2012) Terminated TomoTherapy Inc. discovered an issue on the Hi-Art system during the course of ongoing testing. It has been identified that treatment plans involving very small structures (volume <2cc), reading in a high dose gradient region, may exhibit inaccuracies with the DVH curve and the Dose Statistics for the small structures. This error can only happen if planning was performed using very small voxels
Full distribution text
- 22 Feb 2008 Z-1615-2008 Unclassified (pre-2012) Terminated Registration adjustment field numbers may appear on printed pages of the Register Tab as nonsense characters or values different than the numbers shown on the Operator Station (OS) display. Setup will move the couch to the correct offset position. If the couch is manually adjusted to the incorrect value, the patient may not be the correct position during treatment.
Full distribution text
- 18 Sep 2008 Z-0339-2009 Unclassified (pre-2012) Terminated TomoTherapy Inc. discovered an issue on the Hi-Art system during the course of ongoing testing. It has been identified that treatment plans involving very small structures (volume <2cc), reading in a high dose gradient region, may exhibit inaccuracies with the DVH curve and the Dose Statistics for the small structures. This error can only happen if planning was performed using very small voxels
Frequently asked questions
How many devices does Tomotherapy Incorporated have in the registry?
The registry currently lists 6 devices for Tomotherapy Incorporated. Devices recorded under merged duplicate company entries are counted once.
Which registries list Tomotherapy Incorporated's devices?
Among its 6 most recently registered devices, Tomotherapy Incorporated has devices listed in FDA 510(k) (6 devices).
What risk classes do Tomotherapy Incorporated's devices fall under?
Among its 6 most recently registered devices, the classes recorded for Tomotherapy Incorporated's devices are Class II. Each device page shows its own class and authorisation.