Modification to Synthes Matrix System
by Synthes Spine 510(k)
Active
510(k) K100952
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK100952 | Date5 Aug 2010 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0226-2015
Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.
-
- Terminated Z-3207-2011
Labeling Correction -- The locking cap was not able to be loosened, resulting in the breakage of the screwdriver or the rod to be cut and the assembly to be removed (screw, cap, and rod). This poses a potential risk of significant prolongation of OR time.
Specification
- 510(k) Number
- FDA Product Code
- NKB
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3070
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent