Monaco Rtp System
by Computerized Medical Systems, Inc. 2 variants
510(k) K071938
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK071938 | Date1 Oct 2007 | StatusActive |
Recalls
-
- Terminated Z-2598-2011
Monaco: When the isocenter location is edited in the IMRT Activity screen, the shift coordinates are not updated in the IMRT Plan report.
-
- Terminated Z-2561-2011
Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.
-
- Terminated Z-2569-2011
Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.
-
- Terminated Z-2564-2011
Monaco: due to a software defect, the Radiological Depth (cm) and SSD to DRP (cm) parameters are reported incorrectly on the Beam Summary Report. This occurs for all Monaco IMRT plans.
-
- Terminated Z-2560-2011
Monaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may only use the pencil beam algorithm for final computation of dose even though Monte Carlo was selected. Other than specific output to the console window, there will be no other in
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 27 Sep 2026
The Monaco Rtp System is a radiation therapy treatment planning software device. It is designed to accept patient imaging data and linear accelerator dosimetry data to create intensity modulated radiation therapy plans using multileaf collimators.
This system is used by dosimetrists and medical physicists to contour target volumes and critical structures, as well as to define radiation beam parameters such as number, position, energy, and modifiers. The device is classified as a Class 2 medical device and received FDA 510(k) clearance on October 1, 2007.
Other devices from Computerized Medical Systems, Inc.
Questions
AI-generated answers, 27 Sep 2026
What is the primary function of the Monaco RTP System?
The Monaco RTP System is a radiation therapy treatment planning software device. It is designed to accept patient imaging data and linear accelerator dosimetry data to create intensity modulated radiation therapy plans using multileaf collimators.
Who is the intended user for this software?
This system is intended for use by dosimetrists and medical physicists. These professionals use the software to contour target volumes and critical structures, as well as to define radiation beam parameters such as number, position, energy, and modifiers.
What is the regulatory status of the Monaco RTP System?
The Monaco RTP System is a Class 2 medical device that received FDA 510(k) clearance on October 1, 2007, under the K-number K071938.