Monaco Rtp System, Vmat Option
by Computerized Medical Systems, Inc.
510(k) K091179
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK091179 | Date16 Jul 2009 | StatusActive |
Recalls
-
- Terminated Z-2598-2011
Monaco: When the isocenter location is edited in the IMRT Activity screen, the shift coordinates are not updated in the IMRT Plan report.
-
- Terminated Z-2561-2011
Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.
-
- Terminated Z-2569-2011
Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.
-
- Terminated Z-2564-2011
Monaco: due to a software defect, the Radiological Depth (cm) and SSD to DRP (cm) parameters are reported incorrectly on the Beam Summary Report. This occurs for all Monaco IMRT plans.
-
- Terminated Z-2560-2011
Monaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may only use the pencil beam algorithm for final computation of dose even though Monte Carlo was selected. Other than specific output to the console window, there will be no other in
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 27 Sep 2026
The Monaco Rtp System, Vmat Option is a radiation therapy treatment planning system. It is designed to assist medical professionals in optimizing treatment plans for cancer patients by enhancing precision in targeting tumors while sparing healthy tissue.
This device is categorized as a Class 2 medical device under regulation number 892.5050. It has received FDA 510(k) clearance as a substantially equivalent system for use in clinical radiation therapy environments.
Other devices from Computerized Medical Systems, Inc.
Questions
AI-generated answers, 27 Sep 2026
What is the primary function of the Monaco RTP System, VMAT Option?
This device is a radiation therapy treatment planning system. It is designed to assist medical professionals in creating optimized treatment plans for cancer patients by increasing the precision of tumor targeting while simultaneously sparing surrounding healthy tissue.
What is the regulatory status of this system?
The Monaco RTP System, VMAT Option has received FDA 510(k) clearance. It is classified as a Class 2 medical device under regulation number 892.5050 and has been determined to be a substantially equivalent system for use in clinical radiation therapy environments.
Which medical specialty is this device associated with?
The system is categorized under the Radiology medical specialty. It is intended for use by clinical professionals within radiation therapy environments to support the planning and delivery of treatment.