Skip to content

Multi-Analyte Detection Systems

by Bio-Rad Laboratories

510(k) K112810

Multi-Analyte Detection Systems is a medical device (FDA Class II; FDA product code NHX, Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)) made by Bio-Rad Laboratories, cleared under FDA 510(k) K112810 on 26 Oct 2011, active as of 9 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 9 Oct 2026.

Regulatory status

as of registry sync 9 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK112810 Date26 Oct 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
27 Sep 2011
FDA decision
26 Oct 2011 (Substantially Equivalent)
FDA review time
29 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K112810 (openFDA), as read on 9 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-1377-2023 FDA Class II initiated 21 Feb 2023

    Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.

    • Ongoing Z-1378-2023 FDA Class II initiated 21 Feb 2023

    Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.

Timeline

3 dated events

  1. Cleared FDA 510(k) K112810
  2. Recall initiated FDA recalls Z-1377-2023
  3. Recall initiated FDA recalls Z-1378-2023

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
NHX
Device class
Class II
Regulation
21 CFR Part 866.5775
Advisory committee
Immunology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code NHX (Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)).

Other devices from Bio-Rad Laboratories