Multi-Analyte Detection Systems
510(k) K112810
Multi-Analyte Detection Systems is a medical device (FDA Class II; FDA product code NHX, Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)) made by Bio-Rad Laboratories, cleared under FDA 510(k) K112810 on 26 Oct 2011, active as of 9 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK112810 | Date26 Oct 2011 | StatusActive |
Official records
- FDA 510(k) database K112810
From the FDA record
- Received by the FDA
- 27 Sep 2011
- FDA decision
- 26 Oct 2011 (Substantially Equivalent)
- FDA review time
- 29 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1377-2023
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.
-
- Ongoing Z-1378-2023
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.
Timeline
- Cleared FDA 510(k) K112810
- Recall initiated FDA recalls Z-1377-2023
- Recall initiated FDA recalls Z-1378-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- NHX
- Device class
- Class II
- Regulation
- 21 CFR Part 866.5775
- Advisory committee
- Immunology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code NHX (Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)).
- IMMUNOSCAN RA ANTI-CCP TEST SYSTEM by Euro-Diagnostica AB
- AXSYM ANTI-CCP REAGENT, STANDARD CALIBRATOR AND CONTROL KITS, MODELS 3L91-20, 3L91-01 AND 3L91-10 by Axis-Shield Diagnostics Limited
- ROCHE ELECSYS ANTI-CCP IMMUNOASSAY, ROCHE ELECSYS PRECICONTROL ANTI-CCP by Roche Diagnostics Corp.
- QUANTA Flash CCP3, QUANTA Flash CCP3 Calibrators, and QUANTA Flash CCP3 by Inova Diagnostics. Inc.
- CCPOINT by Euro-Diagnostica AB
- QUANTA LITE CCP ELISA by Inova Diagnostics. Inc.
- IMMULITE 2000 ANTI-CCP IGG ASSAY by Siemens Healthcare Diagnostics
- ADVIA Centaur Anti-CCP IgG (aCCP) assay including Reagents and Calibrators, Quality Controls and Master Curve Materials. by Axis-Shield Diagnostics Limited
- ELIA CCP, MODEL 14-5515-01; ELIA CCP CONTROL, MODEL 83-1009-01 by Sweden Diagnostics (Us), Inc.
- THERATEST EL-ANTI-CCP/2 by THERATEST LABORATORIES, INC.