Skip to content

113-03-01

by Exactech, Inc.

113-03-01 is an orthopaedic implant impactor (FDA Class II) made by Exactech, Inc., cleared under FDA 510(k) K173583 on 12 Mar 2018 (6 earlier submissions) and listed in FDA GUDID since 19 Mar 2026 (UDI-DI 10885862663221), active as of 23 Oct 2026. 7 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number10885862663221 UDI-DI Date19 Mar 2026ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK173583 Date12 Mar 2018DecidedStatusActive

Official records

From the FDA record

Device description
Femoral Head Impactor
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K173583 decided 12 Mar 2018 Active· +6 more in the status table
  • FDA GUDID UDI-DI 10885862663221 listed 19 Mar 2026 Listed

Recalls

7 recalls · 4 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1166-2025 FDA Class II initiated 31 Dec 2024
    • Ongoing Z-1165-2025 FDA Class II initiated 31 Dec 2024

    Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.

    • Ongoing Z-0577-2024 FDA Class II initiated 21 Nov 2023

    Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

    • Ongoing Z-1727-2022 FDA Class II initiated 11 Aug 2022

    Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

  • 2 recalls, same reason

    • Completed Z-2251-2021 FDA Class II initiated 29 Jun 2021
    • Completed Z-2250-2021 FDA Class II initiated 29 Jun 2021

    Biolox Delta Femoral Head was packaged and labeled as 170-36-00 (36mm +0), but the device was marked as 170-36-03 (36mm +3.5).

    • Terminated Z-0572-2008 FDA Unclassified (pre-2012) initiated 1 Nov 2007

    Mislabeled: A ten piece lot of 140-32-03, 12/14 BIOLOXforte Alumina Femoral Heads 32mm (1069020 - 1069029), from Ceramtec were incorrectly labeled as 140-28-03 12/14 BIOLOXforte Alumina Femoral Heads, 28mm

Timeline

15 dated events

  1. Recall initiated FDA recalls Z-0572-2008
  2. Cleared FDA 510(k) K103012
  3. Cleared FDA 510(k) K173583
  4. Recall initiated FDA recalls Z-2251-2021
  5. Recall initiated FDA recalls Z-2250-2021
  6. Recall initiated FDA recalls Z-1727-2022
  7. Recall initiated FDA recalls Z-0577-2024
  8. Recall initiated FDA recalls Z-1166-2025
  9. Recall initiated FDA recalls Z-1165-2025
  10. Listed FDA GUDID 10885862663221

5 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Generic name
Orthopaedic implant impactor, reusable
Model / Ref / Catalog
113-03-01
Primary DI / UDI-DI
10885862663221
FDA product code
JDI, LWJ, LPH, OQI, LZO, MEH
GMDN term
Orthopaedic implant impactor, reusable
Device class
Class II
Regulation number
888.3350, 888.3360, 888.3358, 888.3353
GUDID distribution status
In Commercial Distribution
Sterilisation method
Moist Heat or Steam Sterilization

Similar devices

Other devices filed under FDA product code JDI (Prosthesis, hip, semi-constrained, metal/polymer, cemented).

Other devices from Exactech, Inc.

See all devices from Exactech, Inc. →

Properties

  • Sterilise before use
  • Prescription use
  • Lot / batch number
Other properties (12, all answered No)