Skip to content

N/A 3550-06

by Medtronic, Inc.

Listed FDA GUDID listing, no authorization on record

Valid until

Not applicable

Database listing, no authorization

FDA recalls

4 ongoing

This device is named in the recall

Documents

None

No authorization documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source GUDID, synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00643169953581 UDI-DI Date14 May 2018ListedStatusListed GUDID: in commercial distribution

Recalls

9 recalls · 4 ongoing · FDA

    • Ongoing Z-0372-2025 FDA Class II initiated 18 Sep 2024

    The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.

    • Ongoing Z-1691-2024 FDA Class II initiated 28 Mar 2024

    inability to reprogram one device

    • Ongoing Z-0780-2024 FDA Class II initiated 19 Dec 2023

    Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to adjust stimulation parameters. In the event of this issue, the patient may receive inadequate or excessive therapy and experience non-optimal pain relief.

    • Ongoing Z-2159-2021 FDA Class II initiated 9 Jul 2021

    Specific PINs of the Vectris SureScan MRI lead kits contain the incorrect lead electrode spacing information printed on the shelf box and the sterile pack labeling.

    • Terminated Z-1926-2016 FDA Class II initiated 3 May 2016

    Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.

    • Terminated Z-1147-2015 FDA Class II initiated 9 Jan 2015

    Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.

    • Terminated Z-1524-2014 FDA Class II initiated 27 Feb 2014

    Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval

    • Terminated Z-0278-2014 FDA Class II initiated 30 Sep 2013

    Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).

    • Terminated Z-1536-06 FDA Unclassified (pre-2012) initiated 2 Aug 2006

    Model 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled 3778 lot V009546 may contain a 3777 standard-spaced lead. The result of this error is an overall increase or decrease of 14 mm in the electrode coverage.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Specification

GUDID

Generic Name
Subcutaneous tunneller, single-use
Model / Reference
3550-06
Primary DI / UDI-DI
00643169953581
PMA Number
FDA Product Code
MHY, LGW, NHL, QRB
GMDN Term
Subcutaneous tunneller, single-use
Device Class
Class III
GUDID distribution status
In Commercial Distribution

Other devices from Medtronic, Inc.

See all devices from Medtronic, Inc. →

Properties

  • Sterile
  • Single Use
  • Prescription Use
  • Lot / Batch Number

Other properties (10)

  • Implantable: No
  • Contains Latex: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No