9733309
9733309 is a neurological stereotactic surgery system (FDA Class II; FDA product code LNH, System, nuclear magnetic resonance imaging) made by Medtronic Navigation Inc., cleared under FDA 510(k) K032541 on 12 Sep 2003 and listed in FDA GUDID since 27 May 2016 (UDI-DI 00613994423955), active as of 8 Nov 2026. 2 FDA recalls are linked to it.
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00613994423955 | Date27 May 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK032541 | Date12 Sep 2003 | StatusActive |
Official records
- AccessGUDID record 00613994423955
- FDA 510(k) database K032541
From the FDA record
- Device description
- PRF 9733309 INTERCHANGEABLE
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K032541 decided 12 Sep 2003 Active
- FDA GUDID UDI-DI 00613994423955 listed 27 May 2016 Listed
Recalls
-
- Terminated Z-2129-2016
Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
-
- Terminated Z-2240-2010
Electronic box on headstrap may overheat, causing burn to patient.
Timeline
- Cleared FDA 510(k) K032541
- Recall initiated FDA recalls Z-2240-2010
- Listed FDA GUDID 00613994423955
- Recall initiated FDA recalls Z-2129-2016
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Neurological stereotactic surgery system
- Model / reference
- 9733309
- Primary DI / UDI-DI
- 00613994423955
- 510(k) Number
- FDA product code
- LNH
- GMDN term
- Neurological stereotactic surgery system
- Device class
- Class II
- Regulation number
- 892.1000
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- MRI safety
- MR Conditional
Similar devices in this category
Other devices filed under Neurological stereotactic surgery system.
- Alignment Bone Anchor Cranial 2.1/3.8 56424 by Brainlab SE
- Drill Guide Depth Control (60-100 mm) 51839-02 by Brainlab SE
- ORTHOPILOT FS633 by Aesculap AG
- Navient 955-NC-NC by Claronav Kolahi
- neuromate tool-holder neuro|mate Standard tool-holder by RENISHAW MAYFIELD
- Leksell Vantage Stereotactic System 1040017 1040032 by Elekta Solutions AB
- Unibody™ Bone Fiducials (10mm) FM-4010-00856 by FHC, INC.
- Unibody™ Bone Fiducials (13mm) FM-4013-00857 by FHC, INC.
- PULSE 5050195 by NuVasive, LLC
- ARVIS IN-14200 by KICO KNEE INNOVATION COMPANY PTY LIMITED