9026-21
by Zimmer, Inc.
9026-21 (00-9026-021-00) is a device of GMDN type metallic femoral head prosthesis (FDA Class II) made by Zimmer, Inc., cleared under FDA 510(k) K953337 on 22 Jan 1996 (1 earlier submission) and listed in FDA GUDID since 24 Sep 2015 (UDI-DI 00889024155978), active as of 23 Oct 2026. 23 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00889024155978 | Date24 Sep 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK953337 | Date22 Jan 1996 | StatusActive |
Official records
- AccessGUDID record 00889024155978
- FDA 510(k) database K953337
- FDA 510(k) database K840643
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K953337 decided 22 Jan 1996 Active
- FDA GUDID UDI-DI 00889024155978 listed 24 Sep 2015 Listed
Recalls
-
- Terminated Z-0722-2018
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
-
- Terminated Z-2580-2017
- Terminated Z-2577-2017
- Terminated Z-2581-2017
- Terminated Z-2579-2017
- Terminated Z-2578-2017
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
-
- Terminated Z-1683-2016
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
-
- Terminated Z-1126-2016
- Terminated Z-1137-2016
- Terminated Z-1127-2016
- Terminated Z-1131-2016
- Terminated Z-1110-2016
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
-
- Terminated Z-0849-2016
LDPE bag containing the implant adheres to the highly polished implant surface.
-
- Terminated Z-1033-2014
- Terminated Z-1034-2014
- Terminated Z-1025-2014
- Terminated Z-1031-2014
- Terminated Z-1036-2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
-
- Terminated Z-0070-2014
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
-
- Terminated Z-2290-2010
- Terminated Z-2291-2010
Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
-
- Terminated Z-0198-2010
Manufacturing material may be present on the surface of the implant.
-
- Terminated Z-1066-05
May be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.
Timeline
- Recall initiated FDA recalls Z-1131-2016
- Recall initiated FDA recalls Z-1110-2016
- Recall initiated FDA recalls Z-0849-2016
- Recall initiated FDA recalls Z-1683-2016
- Recall initiated FDA recalls Z-2580-2017
- Recall initiated FDA recalls Z-2577-2017
- Recall initiated FDA recalls Z-2581-2017
- Recall initiated FDA recalls Z-2579-2017
- Recall initiated FDA recalls Z-2578-2017
- Recall initiated FDA recalls Z-0722-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Metallic femoral head prosthesis
- Model / Ref / Catalog
- 9026-21, 00-9026-021-00
- Primary DI / UDI-DI
- 00889024155978
- FDA product code
- LPH, JDI
- GMDN term
- Metallic femoral head prosthesis
- Device class
- Class II
- Regulation number
- 888.3358, 888.3350
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Metallic femoral head prosthesis.
- Biomet® Hip System 163661 by Biomet Orthopedics, Inc.
- CoCrMo Alloy Femoral Heads 12/14 0524222 by ADLER ORTHO SPA
- TST Bipolar Head 11137002854 by TST RAKOR VE TIBBİ ALETLER SAN.VE TİC.LTD.ŞTİ.
- CoCrMo Alloy Femoral Heads 12/14 0524321 by ADLER ORTHO SPA
- Femoral Head CoCr BO76.1694.000.00 by Lincotek Bologna S.r.l
- CoCr Femoral Ball Head 01.25.088 by Medacta International SA
- CoCr Femoral Ball Head 01.25.034 by Medacta International SA
- Trinity Acetabular Hip System E321.028 by CORIN LTD
- VARIANT, Hlavice CoCrMo (∢12/14) 28 S O01210-2800-12018 by ProSpon, spol. s r.o.
- CoCr Modular Head extra short E321.432 by CORIN LTD