Skip to content

Beta Hip Prosthesis

by Zimmer, Inc.

510(k) K953337

Beta Hip Prosthesis is a medical device (FDA Class II; FDA product code LPH, Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented) made by Zimmer, Inc., cleared under FDA 510(k) K953337 on 22 Jan 1996, active as of 22 Oct 2026. 14 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

15 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK953337 Date22 Jan 1996ClearedStatusActive

Official records

From the FDA record

Received by the FDA
17 Jul 1995
FDA decision
22 Jan 1996 (Substantially Equivalent)
FDA review time
189 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K953337 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

15 recalls · FDA

    • Terminated Z-0722-2018 FDA Class II initiated 1 Aug 2017

    The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.

  • 5 recalls, same reason

    • Terminated Z-2580-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2578-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2581-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2579-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2577-2017 FDA Class II initiated 28 Feb 2017

    Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.

    • Terminated Z-1127-2016 FDA Class II initiated 11 Jan 2016

    As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

    • Terminated Z-0849-2016 FDA Class II initiated 11 Jan 2016

    LDPE bag containing the implant adheres to the highly polished implant surface.

  • 2 recalls, same reason

    • Terminated Z-1034-2014 FDA Class II initiated 11 Oct 2013
    • Terminated Z-1031-2014 FDA Class II initiated 11 Oct 2013

    Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

    • Terminated Z-0070-2014 FDA Class II initiated 30 Aug 2013

    The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.

  • 2 recalls, same reason

    • Terminated Z-2290-2010 FDA Unclassified (pre-2012) initiated 5 Jul 2010
    • Terminated Z-2291-2010 FDA Unclassified (pre-2012) initiated 5 Jul 2010

    Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.

    • Terminated Z-0198-2010 FDA Unclassified (pre-2012) initiated 3 Aug 2009

    Manufacturing material may be present on the surface of the implant.

    • Terminated Z-1066-05 FDA Unclassified (pre-2012) initiated 15 Apr 2005 on shared 510(k) K953337

    May be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

16 dated events

  1. Recall initiated FDA recalls Z-1034-2014
  2. Recall initiated FDA recalls Z-1031-2014
  3. Recall initiated FDA recalls Z-1127-2016
  4. Recall initiated FDA recalls Z-0849-2016
  5. Recall initiated FDA recalls Z-2580-2017
  6. Recall initiated FDA recalls Z-2578-2017
  7. Recall initiated FDA recalls Z-2581-2017
  8. Recall initiated FDA recalls Z-2579-2017
  9. Recall initiated FDA recalls Z-2577-2017
  10. Recall initiated FDA recalls Z-0722-2018

6 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
LPH
Device class
Class II
Regulation
21 CFR Part 888.3358
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Devices cleared under K953337

130 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K953337
UDI-DISize and differences
32-9026-38
Zimmer, Inc.
32-9026-038-00 / 32-9026-3800889024365247 —On the market
32-9026-48-45
Zimmer, Inc.
32-9026-048-45 / 32-9026-48-4500889024365308 —On the market
32-9026-49-35
Zimmer, Inc.
32-9026-049-35 / 32-9026-49-3500889024365322 —On the market
9026-21
Zimmer, Inc.
00-9026-021-00 / 9026-2100889024155978 —On the market
9026-22-10
Zimmer, Inc.
00-9026-022-10 / 9026-22-1000889024155985 —On the market
VerSys® 7841-11
Zimmer, Inc.
7841-1100889024136007 —On the market
VerSys® 7841-12-30
Zimmer, Inc.
7841-12-3000889024136069 —On the market
VerSys® 7841-14
Zimmer, Inc.
7841-1400889024136120 —On the market
VerSys® 7841-14-30
Zimmer, Inc.
7841-14-3000889024136144 —On the market
VerSys® 7841-15-30
Zimmer, Inc.
7841-15-3000889024136182 —On the market
VerSys® 7841-16-30
Zimmer, Inc.
7841-16-3000889024136229 —On the market
VerSys® 7841-18
Zimmer, Inc.
7841-1800889024136281 —On the market
VerSys® 7843-11-06
Zimmer, Inc.
00-7843-011-06 / 7843-11-0600889024136403 —On the market
VerSys® 7843-12-08
Zimmer, Inc.
00-7843-012-08 / 7843-12-0800889024136472 —On the market
VerSys® 7843-13-06
Zimmer, Inc.
00-7843-013-06 / 7843-13-0600889024136533 —On the market
VerSys® 7843-13-08
Zimmer, Inc.
00-7843-013-08 / 7843-13-0800889024136540 —On the market
VerSys® 7843-13-36
Zimmer, Inc.
00-7843-013-36 / 7843-13-3600889024136588 —On the market
VerSys® 7843-14-08
Zimmer, Inc.
00-7843-014-08 / 7843-14-0800889024136632 —On the market
VerSys® 7843-16-26
Zimmer, Inc.
00-7843-016-26 / 7843-16-2600889024136823 —On the market
VerSys® 7843-16-36
Zimmer, Inc.
00-7843-016-36 / 7843-16-3600889024136830 —On the market
VerSys® 7843-16-56
Zimmer, Inc.
00-7843-016-56 / 7843-16-5600889024136854 —On the market
VerSys® 7843-17-08
Zimmer, Inc.
00-7843-017-08 / 7843-17-0800889024136892 —On the market
VerSys® 7843-17-26
Zimmer, Inc.
00-7843-017-26 / 7843-17-2600889024136908 —On the market
VerSys® 7843-17-56
Zimmer, Inc.
00-7843-017-56 / 7843-17-5600889024136939 —On the market
VerSys® 7843-18-06
Zimmer, Inc.
00-7843-018-06 / 7843-18-0600889024136946 —On the market
VerSys® 7843-18-36
Zimmer, Inc.
00-7843-018-36 / 7843-18-3600889024136984 —On the market
VerSys® 7843-19-36
Zimmer, Inc.
00-7843-019-36 / 7843-19-3600889024137066 —On the market
VerSys® 7843-20-36
Zimmer, Inc.
00-7843-020-36 / 7843-20-3600889024137141 —On the market
VerSys® 7845-12-20
Zimmer, Inc.
7845-12-2000889024137394 —On the market
VerSys® 7845-12-30
Zimmer, Inc.
7845-12-3000889024137400 —On the market
VerSys® 7845-12-50
Zimmer, Inc.
7845-12-5000889024137417 —On the market
VerSys® 7845-13-20
Zimmer, Inc.
7845-13-2000889024137431 —On the market
VerSys® 7845-13-50
Zimmer, Inc.
7845-13-5000889024137455 —On the market
VerSys® 7845-14-20
Zimmer, Inc.
7845-14-2000889024137479 —On the market
VerSys® 7845-14-50
Zimmer, Inc.
7845-14-5000889024137493 —On the market
VerSys® 7845-15-20
Zimmer, Inc.
7845-15-2000889024137516 —On the market
VerSys® 7845-15-30
Zimmer, Inc.
7845-15-3000889024137523 —On the market
VerSys® 7845-16-30
Zimmer, Inc.
7845-16-3000889024137561 —On the market
VerSys® 7845-16-50
Zimmer, Inc.
7845-16-5000889024137578 —On the market
VerSys® 7845-18-20
Zimmer, Inc.
7845-18-2000889024137639 —On the market
VerSys® 7845-19-30
Zimmer, Inc.
7845-19-3000889024137684 —On the market
VerSys® 8018-22-20
Zimmer, Inc.
00-8018-022-20 / 8018-22-2000889024144545 —On the market
VerSys® 8018-22-30
Zimmer, Inc.
00-8018-022-30 / 8018-22-3000889024144552 —On the market
VerSys® 8018-28-14
Zimmer, Inc.
00-8018-028-14 / 8018-28-1400889024144651 —On the market
VerSys® 8018-32-03
Zimmer, Inc.
00-8018-032-03 / 8018-32-0300889024144682 —On the market
VerSys® 8018-32-14
Zimmer, Inc.
00-8018-032-14 / 8018-32-1400889024144705 —On the market
VerSys® 8018-36-01
Zimmer, Inc.
00-8018-036-01 / 8018-36-0100889024144712 —On the market
VerSys® 8018-36-04
Zimmer, Inc.
00-8018-036-04 / 8018-36-0400889024144743 —On the market
VerSys® 8018-36-05
Zimmer, Inc.
00-8018-036-05 / 8018-36-0500889024144750 —On the market
VerSys® 8018-40-04
Zimmer, Inc.
00-8018-040-04 / 8018-40-0400889024144798 —On the market

Similar devices

Other devices filed under FDA product code LPH (Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented).

Other devices from Zimmer, Inc.

See all devices from Zimmer, Inc. →