Beta Hip Prosthesis
by Zimmer, Inc.
510(k) K953337
Beta Hip Prosthesis is a medical device (FDA Class II; FDA product code LPH, Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented) made by Zimmer, Inc., cleared under FDA 510(k) K953337 on 22 Jan 1996, active as of 22 Oct 2026. 14 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK953337 | Date22 Jan 1996 | StatusActive |
Official records
- FDA 510(k) database K953337
From the FDA record
- Received by the FDA
- 17 Jul 1995
- FDA decision
- 22 Jan 1996 (Substantially Equivalent)
- FDA review time
- 189 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-0722-2018
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
-
- Terminated Z-2580-2017
- Terminated Z-2578-2017
- Terminated Z-2581-2017
- Terminated Z-2579-2017
- Terminated Z-2577-2017
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
-
- Terminated Z-1127-2016
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
-
- Terminated Z-0849-2016
LDPE bag containing the implant adheres to the highly polished implant surface.
-
- Terminated Z-1034-2014
- Terminated Z-1031-2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
-
- Terminated Z-0070-2014
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
-
- Terminated Z-2290-2010
- Terminated Z-2291-2010
Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
-
- Terminated Z-0198-2010
Manufacturing material may be present on the surface of the implant.
-
- Terminated Z-1066-05
May be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.
Timeline
- Recall initiated FDA recalls Z-1034-2014
- Recall initiated FDA recalls Z-1031-2014
- Recall initiated FDA recalls Z-1127-2016
- Recall initiated FDA recalls Z-0849-2016
- Recall initiated FDA recalls Z-2580-2017
- Recall initiated FDA recalls Z-2578-2017
- Recall initiated FDA recalls Z-2581-2017
- Recall initiated FDA recalls Z-2579-2017
- Recall initiated FDA recalls Z-2577-2017
- Recall initiated FDA recalls Z-0722-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- LPH
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3358
- Advisory committee
- Orthopedic
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K953337
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| 32-9026-38 | 32-9026-038-00 / 32-9026-38 | 00889024365247 | — | On the market |
| 32-9026-48-45 | 32-9026-048-45 / 32-9026-48-45 | 00889024365308 | — | On the market |
| 32-9026-49-35 | 32-9026-049-35 / 32-9026-49-35 | 00889024365322 | — | On the market |
| 9026-21 | 00-9026-021-00 / 9026-21 | 00889024155978 | — | On the market |
| 9026-22-10 | 00-9026-022-10 / 9026-22-10 | 00889024155985 | — | On the market |
| VerSys® 7841-11 | 7841-11 | 00889024136007 | — | On the market |
| VerSys® 7841-12-30 | 7841-12-30 | 00889024136069 | — | On the market |
| VerSys® 7841-14 | 7841-14 | 00889024136120 | — | On the market |
| VerSys® 7841-14-30 | 7841-14-30 | 00889024136144 | — | On the market |
| VerSys® 7841-15-30 | 7841-15-30 | 00889024136182 | — | On the market |
| VerSys® 7841-16-30 | 7841-16-30 | 00889024136229 | — | On the market |
| VerSys® 7841-18 | 7841-18 | 00889024136281 | — | On the market |
| VerSys® 7843-11-06 | 00-7843-011-06 / 7843-11-06 | 00889024136403 | — | On the market |
| VerSys® 7843-12-08 | 00-7843-012-08 / 7843-12-08 | 00889024136472 | — | On the market |
| VerSys® 7843-13-06 | 00-7843-013-06 / 7843-13-06 | 00889024136533 | — | On the market |
| VerSys® 7843-13-08 | 00-7843-013-08 / 7843-13-08 | 00889024136540 | — | On the market |
| VerSys® 7843-13-36 | 00-7843-013-36 / 7843-13-36 | 00889024136588 | — | On the market |
| VerSys® 7843-14-08 | 00-7843-014-08 / 7843-14-08 | 00889024136632 | — | On the market |
| VerSys® 7843-16-26 | 00-7843-016-26 / 7843-16-26 | 00889024136823 | — | On the market |
| VerSys® 7843-16-36 | 00-7843-016-36 / 7843-16-36 | 00889024136830 | — | On the market |
| VerSys® 7843-16-56 | 00-7843-016-56 / 7843-16-56 | 00889024136854 | — | On the market |
| VerSys® 7843-17-08 | 00-7843-017-08 / 7843-17-08 | 00889024136892 | — | On the market |
| VerSys® 7843-17-26 | 00-7843-017-26 / 7843-17-26 | 00889024136908 | — | On the market |
| VerSys® 7843-17-56 | 00-7843-017-56 / 7843-17-56 | 00889024136939 | — | On the market |
| VerSys® 7843-18-06 | 00-7843-018-06 / 7843-18-06 | 00889024136946 | — | On the market |
| VerSys® 7843-18-36 | 00-7843-018-36 / 7843-18-36 | 00889024136984 | — | On the market |
| VerSys® 7843-19-36 | 00-7843-019-36 / 7843-19-36 | 00889024137066 | — | On the market |
| VerSys® 7843-20-36 | 00-7843-020-36 / 7843-20-36 | 00889024137141 | — | On the market |
| VerSys® 7845-12-20 | 7845-12-20 | 00889024137394 | — | On the market |
| VerSys® 7845-12-30 | 7845-12-30 | 00889024137400 | — | On the market |
| VerSys® 7845-12-50 | 7845-12-50 | 00889024137417 | — | On the market |
| VerSys® 7845-13-20 | 7845-13-20 | 00889024137431 | — | On the market |
| VerSys® 7845-13-50 | 7845-13-50 | 00889024137455 | — | On the market |
| VerSys® 7845-14-20 | 7845-14-20 | 00889024137479 | — | On the market |
| VerSys® 7845-14-50 | 7845-14-50 | 00889024137493 | — | On the market |
| VerSys® 7845-15-20 | 7845-15-20 | 00889024137516 | — | On the market |
| VerSys® 7845-15-30 | 7845-15-30 | 00889024137523 | — | On the market |
| VerSys® 7845-16-30 | 7845-16-30 | 00889024137561 | — | On the market |
| VerSys® 7845-16-50 | 7845-16-50 | 00889024137578 | — | On the market |
| VerSys® 7845-18-20 | 7845-18-20 | 00889024137639 | — | On the market |
| VerSys® 7845-19-30 | 7845-19-30 | 00889024137684 | — | On the market |
| VerSys® 8018-22-20 | 00-8018-022-20 / 8018-22-20 | 00889024144545 | — | On the market |
| VerSys® 8018-22-30 | 00-8018-022-30 / 8018-22-30 | 00889024144552 | — | On the market |
| VerSys® 8018-28-14 | 00-8018-028-14 / 8018-28-14 | 00889024144651 | — | On the market |
| VerSys® 8018-32-03 | 00-8018-032-03 / 8018-32-03 | 00889024144682 | — | On the market |
| VerSys® 8018-32-14 | 00-8018-032-14 / 8018-32-14 | 00889024144705 | — | On the market |
| VerSys® 8018-36-01 | 00-8018-036-01 / 8018-36-01 | 00889024144712 | — | On the market |
| VerSys® 8018-36-04 | 00-8018-036-04 / 8018-36-04 | 00889024144743 | — | On the market |
| VerSys® 8018-36-05 | 00-8018-036-05 / 8018-36-05 | 00889024144750 | — | On the market |
| VerSys® 8018-40-04 | 00-8018-040-04 / 8018-40-04 | 00889024144798 | — | On the market |
Similar devices
Other devices filed under FDA product code LPH (Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented).
- G7® Acetabular System by Zimmer Biomet
- Precision Casting Solutions Total Hip System by Precision Casting Solution, LLC
- OSTEOLOCK ACETABULAR SHELL SYSTEM by Howmedica Corp.
- Delta Alumina Ceramic Heads Morse Taper 12/14 Femoral Head 0514181 by ADLER ORTHO SPA
- G7® Freedom® 010002607 by Biomet Orthopedics, Inc.
- G7® 010002711 by Biomet Orthopedics, Inc.
- SMITH & NEPHEW OXINIUM DH FEMORAL HEADS by Smith & Nephew, Inc.
- BIOLOX DELTA 5010.42.361 by Limacorporate
- Mpact Acetabular System 01.32.158SH by Medacta International SA
- POLARCUP DUAL MOBILITY SYSTEM by Smith & Nephew, Inc.