11141-000156 is a primary battery (FDA Class III; FDA product code MKJ, Automated external defibrillators (non-wearable)) made by PHYSIO-CONTROL, Inc., approved under FDA PMA P160026 on 2 Jul 2018 and listed in FDA GUDID since 10 Nov 2014 (UDI-DI 00883873919477), active as of 23 Oct 2026. 17 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

17 ongoing

This device is named in the recall

Documents

PMA P160026

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA PMA from GUDID NumberP160026 Date2 Jul 2018IssuedStatusActive
United StatesGUDID listing Number00883873919477 UDI-DI Date10 Nov 2014ListedStatusListed GUDID: in commercial distribution

Official records

From the FDA record

Device description
LP1000 LITHIUM METAL NONRECHARGEABLE BATTERY
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA PMA P160026 decided 2 Jul 2018 Active
  • FDA GUDID UDI-DI 00883873919477 listed 10 Nov 2014 Listed

Recalls

17 recalls · 17 ongoing · FDA

    • Ongoing Z-2065-2026 FDA Class II initiated 13 Apr 2026

    Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

  • 2 recalls, same reason

    • Ongoing Z-2497-2025 FDA Class II initiated 22 Jul 2025
    • Ongoing Z-2498-2025 FDA Class II initiated 22 Jul 2025

    Due to required inspections not being performed on products/units that have gone through servicing.

    • Ongoing Z-1232-2025 FDA Class II initiated 3 Feb 2025

    Incorrect keypad was incorrectly installed onto a defibrillator/monitor.

    • Ongoing Z-1184-2025 FDA Class II initiated 21 Jan 2025

    Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.

    • Ongoing Z-0230-2025 FDA Class II initiated 30 Aug 2024

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    • Ongoing Z-0567-2024 FDA Class II initiated 6 Nov 2023

    AC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.

  • 10 recalls, same reason

    • Ongoing Z-0296-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0303-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0305-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0302-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0298-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0301-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0299-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0297-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0300-2024 FDA Class II initiated 29 Sep 2023
    • Ongoing Z-0304-2024 FDA Class II initiated 29 Sep 2023

    Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

Timeline

19 dated events

  1. Recall initiated FDA recalls Z-0297-2024
  2. Recall initiated FDA recalls Z-0300-2024
  3. Recall initiated FDA recalls Z-0304-2024
  4. Recall initiated FDA recalls Z-0567-2024
  5. Recall initiated FDA recalls Z-0230-2025
  6. Recall initiated FDA recalls Z-1184-2025
  7. Recall initiated FDA recalls Z-1232-2025
  8. Recall initiated FDA recalls Z-2497-2025
  9. Recall initiated FDA recalls Z-2498-2025
  10. Recall initiated FDA recalls Z-2065-2026

9 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

GUDID

Generic name
Primary battery
Model / Ref / Catalog
11141, 11141-000156
Primary DI / UDI-DI
00883873919477
PMA number
FDA product code
MKJ
GMDN term
Primary battery
Device class
Class III
Regulation number
870.5310
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code MKJ (Automated external defibrillators (non-wearable)).

Other devices from PHYSIO-CONTROL, Inc.

See all devices from PHYSIO-CONTROL, Inc. →

Properties

  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (12, all answered No)