2326
Part of a family of 13 devices: see the family page
Ref 2326
2326 is a neural-tissue electrical stimulation lead adaptor (EU MDR Class III; FDA Class III; FDA product code LGW, Stimulator, Spinal-Cord, Totally Implanted for Pain Relief) made by St. Jude Medical, Inc., registered in EUDAMED on 25 Aug 2026 (Basic UDI-DI 5415067ACC000254) with CE/MDR certificate MDR 779805 valid until 11 Dec 2028, approved under FDA PMA P010032 on 21 Nov 2001 and listed in FDA GUDID since 22 Apr 2015 (UDI-DI 05415067016140), active as of 24 Oct 2026. 14 FDA recalls are linked to it.
Active
Valid until
11 Dec 2028
FDA recalls
8 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number5415067ACC000254 05415067016140 | Date25 Aug 2026 | StatusActive |
| United States | GUDID listing GUDID | Number05415067016140 | Date22 Apr 2015 | StatusListed |
| United States | FDA PMAAPPR GUDID | NumberP010032 | Date21 Nov 2001 | StatusActive |
Official records
- EUDAMED public site 5415067ACC000254
- AccessGUDID record 05415067016140
- FDA PMA database P010032
From the FDA record
- Device description
- 8-channel adapter, B, 60cm
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- EUDAMED Basic UDI-DI 5415067ACC000254 registered 25 Aug 2026 Active
- FDA PMA P010032 decided 21 Nov 2001 Active
- FDA GUDID UDI-DI 05415067016140 listed 22 Apr 2015 Listed
Certificates
- MDR 779805 Active
Technical documentation EU MDR
- Certifier
- BSI Group The Netherlands B.V. 2797
- Issued
- 25 Jun 2026
- Expires
- 11 Dec 2028
- Reason
- Other
Recalls
-
- Ongoing Z-0458-2026
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
-
- Ongoing Z-2505-2023
- Ongoing Z-2506-2023
- Ongoing Z-2507-2023
- Ongoing Z-2508-2023
- Ongoing Z-2509-2023
- Ongoing Z-2510-2023
- Ongoing Z-2511-2023
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
-
- Terminated Z-2691-2017
- Terminated Z-2692-2017
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
-
- Terminated Z-1785-2013
There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the
-
- Terminated Z-0025-2013
Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
-
- Terminated Z-1738-2013
St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica
-
- Terminated Z-3091-2011
Product programmed with unapproved software was distributed.
Timeline
- Recall initiated FDA recalls Z-2506-2023
- Recall initiated FDA recalls Z-2507-2023
- Recall initiated FDA recalls Z-2508-2023
- Recall initiated FDA recalls Z-2509-2023
- Recall initiated FDA recalls Z-2510-2023
- Recall initiated FDA recalls Z-2511-2023
- Recall initiated FDA recalls Z-0458-2026
- Certificate issued EUDAMED MDR 779805
- Registered EUDAMED 5415067ACC000254
- Certificate expires EUDAMED MDR 779805
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic name
- Neural-tissue electrical stimulation lead adaptor
- Device name
- Neuromodulation Accessories
- Model / Ref / Catalog
- 2326
- Primary DI / UDI-DI
- 05415067016140
- Basic UDI-DI
- 5415067ACC000254
- PMA number
- FDA product code
- LGW
- EMDN code
- J0280 Implantable neurostimulators - accessories
- GMDN term
- Neural-tissue electrical stimulation lead adaptor
- Device class
- Class III
- EU registration type
- Standard device
- EU market of registration
- Belgium
- GUDID distribution status
- In Commercial Distribution
- Size
- Length — 60 Centimeter
Similar devices in this category
Other devices filed under IMPLANTABLE NEUROSTIMULATORS.
- Precision ™ SC-5212 by Boston Scientific Neuromodulation
- Precision™ SC-4400 by Boston Scientific Neuromodulation
- Evoke® 12C Trial Percutaneous Lead Kit - 60cm 1016 by Saluda Medical Pty Ltd
- VNS Therapy® 105 by LivaNova USA, Inc.
- SCS Freedom-4A Trial Neurostimulator Kit Channel A FR4A-TRL-A0 by Curonix LLC
- VNS Therapy® 103 by LivaNova USA, Inc.
- remede 1004A-F by ZOLL Respicardia, Inc.
- Evoke® 12C Trial Percutaneous Lead Kit - 90cm 1017 by Saluda Medical Pty Ltd
- Barostim™ Programmer System 100067-201 by Cvrx, Inc.
- remede 1004A by ZOLL Respicardia, Inc.