VNS Therapy®

by LivaNova USA, Inc. 8 variants

Ref 105Model Model 105

VNS Therapy (105) is a medical device (EU AIMDD; FDA Class III) made by LivaNova USA, Inc., registered in EUDAMED on 6 May 2026 (Basic UDI-DI B-05425025750054), approved under FDA PMA P970003 on 16 Jul 1997 and listed in FDA GUDID since 17 Sep 2014 (UDI-DI 05425025750054), active as of 22 Oct 2026. 30 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE AIMDD PMA

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

PMA P970003

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/AIMDD EUDAMED NumberB-05425025750054 Basic UDI-DI 05425025750054UDI-DIDate6 May 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number05425025750054 UDI-DI Date17 Sep 2014ListedStatusListed GUDID: in commercial distribution
United StatesFDA PMAAPPR GUDID from GUDID Approved NumberP970003 Date16 Jul 1997DecidedStatusActive

Official records

From the FDA record

Device description
Implantable Pulse Generator
MRI safety
MR Conditional
Natural rubber latex
Labelled as not made with natural rubber latex
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 30 May 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI B-05425025750054 registered 6 May 2026 Active
  • FDA PMA P970003 decided 16 Jul 1997 Active
  • FDA GUDID UDI-DI 05425025750054 listed 17 Sep 2014 Listed

Recalls

30 recalls · 6 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1067-2025 FDA Class II initiated 3 Jan 2025
    • Ongoing Z-1068-2025 FDA Class II initiated 3 Jan 2025

    Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.

    • Ongoing Z-1072-2025 FDA Class II initiated 12 Dec 2024

    Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

    • Ongoing Z-1563-2020 FDA Class II initiated 4 Feb 2020

    False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.

    • Ongoing Z-0346-2020 FDA Class II initiated 16 Nov 2018

    Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during diagnostic testing. As a result, normal impedance ranges for the affected VNS generator have shifted relative to the existing thresholds of 600 - 5300 Ohms defined in labeling and as present in the programming software.

    • Ongoing Z-0663-2020 FDA Class II initiated 16 Nov 2018

    This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.

    • Terminated Z-0532-2022 FDA Class II initiated 2 Dec 2021

    A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

    • Terminated Z-0462-2021 FDA Class II initiated 28 Sep 2020

    Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.

    • Terminated Z-0561-2020 FDA Class I initiated 22 Aug 2019

    Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.

    • Terminated Z-1118-2020 FDA Class II initiated 13 Aug 2019

    Firm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.

  • 2 recalls, same reason

    • Terminated Z-0868-2020 FDA Class II initiated 16 Nov 2018
    • Terminated Z-0869-2020 FDA Class II initiated 16 Nov 2018

    This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.

    • Terminated Z-2572-2018 FDA Class II initiated 24 Jan 2018

    Unintended warning message displayed on generators programmed with a Model 3000 v.1.0.2.2 programmer.

    • Terminated Z-0555-2018 FDA Class II initiated 2 Dec 2017

    Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.

    • Terminated Z-2255-2018 FDA Class II initiated 13 Nov 2017

    Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

    • Terminated Z-0659-2016 FDA Class II initiated 18 Dec 2015

    Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.

    • Terminated Z-0645-2016 FDA Class II initiated 18 Nov 2015

    Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.

    • Terminated Z-0280-2016 FDA Class II initiated 7 Oct 2015

    Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.

    • Terminated Z-0858-2015 FDA Class II initiated 20 Nov 2014 UDI-DI 05425025750054 listed

    The recalled product was distributed with an incorrect serial number printed on the device's label.

    • Terminated Z-0210-2012 FDA Unclassified (pre-2012) initiated 12 Aug 2011

    The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.

  • 2 recalls, same reason

    • Terminated Z-0005-2012 FDA Unclassified (pre-2012) initiated 6 May 2011
    • Terminated Z-0006-2012 FDA Unclassified (pre-2012) initiated 6 May 2011

    An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.

    • Terminated Z-1591-2010 FDA Unclassified (pre-2012) initiated 6 Apr 2010

    Battery life projection is inaccurate.

    • Terminated Z-0994-2009 FDA Unclassified (pre-2012) initiated 7 Feb 2008

    Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.

    • Terminated Z-0995-2009 FDA Unclassified (pre-2012) initiated 7 Feb 2008

    Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.

  • 3 recalls, same reason

    • Terminated Z-0171-2008 FDA Unclassified (pre-2012) initiated 2 Aug 2007
    • Terminated Z-0172-2008 FDA Unclassified (pre-2012) initiated 2 Aug 2007
    • Terminated Z-0173-2008 FDA Unclassified (pre-2012) initiated 2 Aug 2007

    Dissolution/Fractures to the leads of the VNS Therapy System

  • 2 recalls, same reason

    • Terminated Z-0341-2007 FDA Unclassified (pre-2012) initiated 17 Oct 2006
    • Terminated Z-0342-2007 FDA Unclassified (pre-2012) initiated 17 Oct 2006

    During programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician

Timeline

33 dated events

  1. Recall initiated FDA recalls Z-0869-2020
  2. Recall initiated FDA recalls Z-1118-2020
  3. Recall initiated FDA recalls Z-0561-2020
  4. Recall initiated FDA recalls Z-1563-2020
  5. Recall initiated FDA recalls Z-0462-2021
  6. Recall initiated FDA recalls Z-0532-2022
  7. Recall initiated FDA recalls Z-1072-2025
  8. Recall initiated FDA recalls Z-1067-2025
  9. Recall initiated FDA recalls Z-1068-2025
  10. Registered EUDAMED B-05425025750054

23 earlier events are in the sections above.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/AIMDD Active
  • United States FDA PMA Active

Specification

EUDAMED · GUDID

Trade name
VNS Therapy® EUDAMED
VNS Therapy® AspireHC® Model 105 Generator GUDID
Generic name
Antiseizure/psychiatric-therapy vagus nerve implantable electrical stimulation system
Device name
VNS Therapy® AspireHC® Generator
Model / reference
Model 105, 105
Primary DI / UDI-DI
05425025750054
Basic UDI-DI
B-05425025750054
PMA number
FDA product code
MUZ, LYJ
EMDN code
J020301, J020399
GMDN term
Antiseizure/psychiatric-therapy vagus nerve implantable electrical stimulation system
Device class
Class AIMDD EUDAMED
Class III GUDID
EU registration type
Legacy device (under a directive)
EU market of registration
Italy
GUDID distribution status
In Commercial Distribution
MRI safety
MR Conditional

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class AIMDD
GUDID
Class III

Device class is stated on each registry's own scale: Class AIMDD in the EU (EUDAMED), Class III in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Manufacturer details

EUDAMED SRN
US-MF-000038770
Authorised representative
LivaNova Belgium NV BE-AR-000014035

Similar devices in this category

Other devices filed under IMPLANTABLE NEUROSTIMULATORS.

Other devices from LivaNova USA, Inc.

  • Perfusion Pack

    Class II
  • Perfusion Pack

    Class II
  • Perfusion Pack

    Class II
  • Perfusion Pack

    Class II

See all devices from LivaNova USA, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Serial number
  • Expiration date
  • Manufacturing date
Other properties (16, all answered No)