- Ongoing Z-2727-2026
- Ongoing Z-2726-2026
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
-
- Ongoing Z-0811-2025
- Ongoing Z-0812-2025
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
-
- Terminated Z-0770-2016
- Terminated Z-0768-2016
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
-
- Terminated Z-1407-2015
- Terminated Z-1406-2015
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Mobile basic diagnostic x-ray system, digital
- Model / Reference
- 5421698
- Primary DI / UDI-DI
- 00840682119658
- 510(k) Number
- FDA Product Code
- IZL
- GMDN Term
- Mobile basic diagnostic x-ray system, digital
- Device Class
- Class II
- Regulation Number
- 892.1720
- GUDID distribution status
- In Commercial Distribution
Other devices from GE Medical Systems, L.L.C.
Properties
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No