Optima XR200amx
Optima XR200amx is a Class II device made by GE Medical Systems, L.L.C., cleared under FDA 510(k) K142383 on 25 Sep 2014 and listed in FDA GUDID since 1 Aug 2016 (UDI-DI 00840682115278), active as of 23 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682115278 | Date1 Aug 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK142383 | Date25 Sep 2014 | StatusActive |
Official records
- AccessGUDID record 00840682115278
- FDA 510(k) database K142383
Across registries
- FDA 510(k) K142383 decided 25 Sep 2014 Active
- FDA GUDID UDI-DI 00840682115278 listed 1 Aug 2016 Listed
Recalls
-
- Ongoing Z-2727-2026
- Ongoing Z-2726-2026
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
-
- Ongoing Z-0811-2025
- Ongoing Z-0812-2025
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
-
- Terminated Z-0770-2016
- Terminated Z-0768-2016
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
-
- Terminated Z-1407-2015
- Terminated Z-1406-2015
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
Timeline
- Cleared FDA 510(k) K142383
- Recall initiated FDA recalls Z-1407-2015
- Recall initiated FDA recalls Z-1406-2015
- Recall initiated FDA recalls Z-0770-2016
- Recall initiated FDA recalls Z-0768-2016
- Listed FDA GUDID 00840682115278
- Recall initiated FDA recalls Z-0811-2025
- Recall initiated FDA recalls Z-0812-2025
- Recall initiated FDA recalls Z-2727-2026
- Recall initiated FDA recalls Z-2726-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Optima
- Generic name
- Portable basic diagnostic x-ray system, digital
- Model / reference
- XR200amx
- Primary DI / UDI-DI
- 00840682115278
- 510(k) Number
- FDA product code
- IZL
- GMDN term
- Portable basic diagnostic x-ray system, digital
- Device class
- Class II
- Regulation number
- 892.1720
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code IZL (System, x-ray, mobile).
- AcuityPDR by Radmedix, LLC
- GENERAL ELECTRIC AMX 3 by Pro-Tronics, Inc.
- 510 BEDSIDE X-RAY SYSTEM by Union Carbide Corp.
- 9170, 1170 by Dynarad Corp.
- REPLACEMENT BATERY FOR GE AMX-4 PORT. X-RAY MACH. by Portable Power Systems, Inc.
- DRGEM TOPAZ-40D by DRGEM Corporation
- DRX-Revolution Nano Mobile X-ray System by CARESTREAM HEALTH, INC.
- BATTERY PACK by Portable Power Systems, Inc.
- SCD-105 AND SCD-125 CAPACITOR DISCHARGE MOBILE X-R by Acoma Medical Imaging, Inc.
- CAP-MOBILE by Philips Medical Systems (Cleveland), Inc.