Na Sc-6360-72
by Boston Scientific Neuromodulation
Listed
Valid until
Not applicable
FDA recalls
3 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number08714729997283 | Date28 Dec 2020 | StatusListed |
Recalls
-
- Ongoing Z-2879-2026
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
-
- Ongoing Z-2404-2026
Sterile anchors my lack sterility assurance.
-
- Ongoing Z-2617-2024
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
-
- Terminated Z-2124-2020
- Terminated Z-2122-2020
- Terminated Z-2119-2020
- Terminated Z-2121-2020
- Terminated Z-2120-2020
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
-
- Terminated Z-2123-2020
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
-
- Terminated Z-1409-2010
Toshiba has determined that certain Sony battery packs installed in Toshiba portable computers can overheat, posing fire hazard to customers. Based on Toshiba's investigation with Sony Corporation, they concluded that the battery cells in specific manufacturing lots could be affected.
-
- Terminated Z-0271-2009
Since April 2004, 27 of 8,769 (0.31%) patients have reported receiving second degree burns in the area of charging and 3 of 8,769 (0.03%) patients have reported receiving third degree burns in the area of charging while using Charger 1 .0. Some of these events were the direct result of the patient disregarding the instructions for use, such as sleeping with the Charger on or placing it directly on
-
- Terminated Z-1441-2012
The recall was initiated because there is a possibility that the OMG Connector-A is incorrectly packaged with an OMG Connector-M Instruction manual.
-
- Terminated Z-0960-2008
Incorrect Data -- Corruption of internal memory component results in an inability for the physician to reprogram the IPG with firmware version prior to Revision 3.02. When this occurs, the IPG will report an error code of "10h0" or "00h0" through the Remote Control. Under this condition, the IPG will cease to log in some data that could be used for informational purposes. ** It should be noted t
-
- Terminated Z-0055-2008
Mislabeled: Product labeled with the incorrect shelf-life information.
-
- Terminated Z-0700-2007
The Precision Implantable Pulse Generator with firmware Revision 3.01 incorrectly reports an error condition for a valid output from its self internal check.
-
- Terminated Z-0299-2007
A small number of unimplanted Precision Linear Leads may have been assembled with incorrect electrode material. At high stimulation levels, the metal may corrode and dissolved metals may enter the patient.
-
- Terminated Z-0396-05
A burn injury was sustained by a patient that did not follow the directions for use and also slept while charging the unit.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Trial spinal cord/peripheral nerve percutaneous analgesic electrical stimulation system pulse generator
- Model / Reference
- SC-6360-72
- Primary DI / UDI-DI
- 08714729997283
- PMA Number
- FDA Product Code
- LGW, QRB
- GMDN Term
- Trial spinal cord/peripheral nerve percutaneous analgesic electrical stimulation system pulse generator
- Device Class
- Class III
- GUDID distribution status
- In Commercial Distribution
Other devices from Boston Scientific Neuromodulation
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No