Na 37022
Listed
Valid until
Not applicable
FDA recalls
11 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00643169513013 | Date20 Nov 2015 | StatusListed |
Recalls
-
- Ongoing Z-2241-2026
- Ongoing Z-2240-2026
A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.
-
- Ongoing Z-0206-2025
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
-
- Ongoing Z-0372-2025
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
-
- Ongoing Z-1691-2024
inability to reprogram one device
-
- Ongoing Z-0780-2024
Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to adjust stimulation parameters. In the event of this issue, the patient may receive inadequate or excessive therapy and experience non-optimal pain relief.
-
- Ongoing Z-1128-2022
The Implantable Neurostimulator (INS) cannot communicate with the clinical programmer and/or the patient programmer system.
-
- Ongoing Z-0418-2022
During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.
-
- Ongoing Z-0201-2022
A software anomaly may occur with the clinician programmer application.
-
- Ongoing Z-2159-2021
Specific PINs of the Vectris SureScan MRI lead kits contain the incorrect lead electrode spacing information printed on the shelf box and the sterile pack labeling.
-
- Ongoing Z-1368-2021
There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.
-
- Completed Z-0217-2024
There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.
-
- Terminated Z-3069-2017
Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.
-
- Terminated Z-2645-2016
Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.
-
- Terminated Z-2602-2016
Medtronic has voluntarily decided to retrieve unused Model 3387S-40 and 3389S-40 DBS leads from three manufacturing lot numbers. During the manufacturing process of a DBS lead component, Medtronic identified the potential for lead insulation damage.
-
- Terminated Z-1926-2016
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.
-
- Terminated Z-1617-2015
Medtronic is providing Important information regarding extension handling during implant procedure after receiving reports of high impedances. This applies to extension Models 37085 and 37086 which can be used with the following implantable Neurostimulators: Activa PC (Model 37601), Activa RC (Model 37612), and the Activa SC (Model 37603).
-
- Terminated Z-1147-2015
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
-
- Terminated Z-1524-2014
Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
-
- Terminated Z-0278-2014
- Terminated Z-0279-2014
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
-
- Terminated Z-1190-2013
There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag
-
- Terminated Z-0186-2009
- Terminated Z-0185-2009
- Terminated Z-0184-2009
- Terminated Z-0187-2009
Medtronic has received reports of DBS" leads being damaged at the proximal connector end of the lead when the lead cap is used in the implant procedure. The lead damage is the result of excessive torsion and / or tensile lading of the proximal connector end of the lead. Data also suggest that excessive torque while tightening or loosening the setscrews may twist the setscrew connector block and d
-
- Terminated Z-1536-06
Model 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled 3778 lot V009546 may contain a 3777 standard-spaced lead. The result of this error is an overall increase or decrease of 14 mm in the electrode coverage.
-
- Terminated Z-0370-06
A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate (SAR) of not more than 0.4 W/kg. The recommended maximum displayed head SAR is now 0.1 W/kg.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Trial spinal cord/peripheral nerve percutaneous analgesic electrical stimulation system pulse generator
- Model / Reference
- 37022
- Primary DI / UDI-DI
- 00643169513013
- FDA Product Code
- OLM, MRU, LGW, MHY, MFR, NHL, MBX, PJS, SGS
- GMDN Term
- Trial spinal cord/peripheral nerve percutaneous analgesic electrical stimulation system pulse generator
- Device Class
- Unclassified
- GUDID distribution status
- In Commercial Distribution
Other devices from Medtronic, Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No