Nexxis OR
Nexxis OR (K9605023) is a medical device (FDA Class II; FDA product code DXJ, Display, Cathode-Ray Tube, Medical) made by Barco NV, cleared under FDA 510(k) K173381 on 22 Feb 2018 and listed in FDA GUDID since 23 Jul 2020 (UDI-DI 05415334028111), active as of 9 Nov 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number05415334028111 | Date23 Jul 2020 | StatusListed |
| United States | FDA 510(k) | NumberK173381 | Date22 Feb 2018 | StatusActive |
Official records
- AccessGUDID record 05415334028111
- FDA 510(k) database K173381
From the FDA record
- Device description
- NEXXIS SYSTEM X670 CM US V2
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K173381 decided 22 Feb 2018 Active
- FDA GUDID UDI-DI 05415334028111 listed 23 Jul 2020 Listed
Recalls
-
- Ongoing Z-1994-2025
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
-
- Ongoing Z-1854-2022
installation issues
-
- Completed Z-2332-2024
We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
Timeline
- Cleared FDA 510(k) K173381
- Listed FDA GUDID 05415334028111
- Recall initiated FDA recalls Z-1854-2022
- Recall initiated FDA recalls Z-2332-2024
- Recall initiated FDA recalls Z-1994-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Operating room audiovisual data/device management system
- Model / Ref / Catalog
- Nexxis OR, K9605023
- Primary DI / UDI-DI
- 05415334028111
- 510(k) Number
- FDA product code
- DXJ
- GMDN term
- Operating room audiovisual data/device management system
- Device class
- Class II
- Regulation number
- 870.2450
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code DXJ (DISPLAY, CATHODE-RAY TUBE, MEDICAL).
- RADIFORCE LARGE MONITOR SYSTEM by EIZO GmbH
- SERVOMED MEDIVISION SCOPE SMV-235 by Litton Medical Electronics
- 620CSR DISPLAY DRIVER by VIDCO, INC
- TLINK DATA MANAGEMENT SYSTEM (DMS) by Terumo Cardiovascular Systems Corporation
- NIVR58-T kit by Barco N.V.
- MEDICAL OSCILLOSCOPE MEMORY SCOPE by Abbott Laboratories
- HP CAREVUE 9000 CLINICAL INFORMATION SYSTEM, MODEL M2331A by Hewlett-Packard Co.
- SPACELABS MEDICAL INTEGRATED CLINICAL DISPLAY by Spacelabs Medical, Inc.
- ZOLL RESCUENET EPCR MODEL RESCUENET EPCR by Zoll Medical Corporation
- ZOLL CODE WRITER by Zoll Medical Corporation