Oec 9800 Plus Digital Mobile Imaging System
510(k) K021049
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK021049 | Date17 Apr 2002 | StatusActive |
Recalls
-
- Ongoing Z-2341-2025
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
-
- Ongoing Z-0706-2025
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
-
- Ongoing Z-2279-2023
- Ongoing Z-2276-2023
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
-
- Terminated Z-2629-2016
A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
-
- Terminated Z-0195-2015
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
-
- Terminated Z-2241-2012
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
-
- Terminated Z-2240-2012
GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
-
- Terminated Z-2262-2012
GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
-
- Terminated Z-1730-2009
GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have sterile seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state.
-
- Terminated Z-1708-2009
- Terminated Z-1707-2009
- Terminated Z-1706-2009
- Terminated Z-1709-2009
GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state. Eleven complaints and 3 MDR's have been filed. No illness or injuries reported.
-
- Terminated Z-1705-2008
Use of existing four-pedal footswitch on a different machine may cause various operational errors.
-
- Terminated Z-2038-2012
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure.
-
- Terminated Z-2026-2012
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
-
- Terminated Z-2125-2012
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and
-
- Terminated Z-0680-2007
- Terminated Z-0681-2007
- Terminated Z-0677-2007
- Terminated Z-0682-2007
During routine service, primary and/or secondary collimators may not have been properly reinstalled.
-
- Terminated Z-0558-2007
- Terminated Z-0556-2007
- Terminated Z-0557-2007
- Terminated Z-0559-2007
During routine service, a cover may have been installed without the required proper lead shielding.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent