Oec 9800 Plus Digital Mobile Imaging System

by Ge Oec Medical Systems

510(k) K021049

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK021049 Date17 Apr 2002ClearedStatusActive

Recalls

26 recalls · 4 ongoing · FDA

    • Ongoing Z-2341-2025 FDA Class II initiated 2 Jul 2025

    GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

    • Ongoing Z-0706-2025 FDA Class II initiated 21 Nov 2024 on shared 510(k) K021049

    Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

  • 2 recalls, same reason

    • Ongoing Z-2279-2023 FDA Class II initiated 15 May 2023 on shared 510(k) K021049
    • Ongoing Z-2276-2023 FDA Class II initiated 15 May 2023

    Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.

    • Terminated Z-2629-2016 FDA Class II initiated 22 Jul 2016

    A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.

    • Terminated Z-0195-2015 FDA Class II initiated 3 Oct 2014

    GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.

    • Terminated Z-2241-2012 FDA Class II initiated 26 Apr 2010 on shared 510(k) K021049

    GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.

    • Terminated Z-2240-2012 FDA Class II initiated 6 Nov 2009 on shared 510(k) K021049

    GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.

    • Terminated Z-2262-2012 FDA Class II initiated 10 Oct 2008 on shared 510(k) K021049

    GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.

    • Terminated Z-1730-2009 FDA Unclassified (pre-2012) initiated 29 Aug 2008 on shared 510(k) K021049

    GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have sterile seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state.

  • 4 recalls, same reason

    • Terminated Z-1708-2009 FDA Unclassified (pre-2012) initiated 15 Aug 2008 on shared 510(k) K021049
    • Terminated Z-1707-2009 FDA Unclassified (pre-2012) initiated 15 Aug 2008 on shared 510(k) K021049
    • Terminated Z-1706-2009 FDA Unclassified (pre-2012) initiated 15 Aug 2008 on shared 510(k) K021049
    • Terminated Z-1709-2009 FDA Unclassified (pre-2012) initiated 15 Aug 2008 on shared 510(k) K021049

    GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state. Eleven complaints and 3 MDR's have been filed. No illness or injuries reported.

    • Terminated Z-1705-2008 FDA Unclassified (pre-2012) initiated 15 Apr 2008 on shared 510(k) K021049

    Use of existing four-pedal footswitch on a different machine may cause various operational errors.

    • Terminated Z-2038-2012 FDA Class I initiated 22 Feb 2008 on shared 510(k) K021049

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure.

    • Terminated Z-2026-2012 FDA Class II initiated 19 Nov 2007 on shared 510(k) K021049

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

    • Terminated Z-2125-2012 FDA Class II initiated 27 Mar 2007 on shared 510(k) K021049

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and

  • 4 recalls, same reason

    • Terminated Z-0680-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007 on shared 510(k) K021049
    • Terminated Z-0681-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007 on shared 510(k) K021049
    • Terminated Z-0677-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007 on shared 510(k) K021049
    • Terminated Z-0682-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007 on shared 510(k) K021049

    During routine service, primary and/or secondary collimators may not have been properly reinstalled.

  • 4 recalls, same reason

    • Terminated Z-0558-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007 on shared 510(k) K021049
    • Terminated Z-0556-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007 on shared 510(k) K021049
    • Terminated Z-0557-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007 on shared 510(k) K021049
    • Terminated Z-0559-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007 on shared 510(k) K021049

    During routine service, a cover may have been installed without the required proper lead shielding.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
OWB
Device Class
Class II
Regulation
21 CFR Part 892.1650
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent