RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK921766 Date23 Apr 1993ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-2323-2017 FDA Class II initiated 5 May 2017

    GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FMT
Device Class
Class II
Regulation
21 CFR Part 880.5130
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 26 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

An infant radiant warmer is a device that provides a controlled source of radiant heat for infants and pediatric patients. The Ohio Infant Warmer System uses a microprocessor-based control system to deliver heat in both manual and servo modes of operation.

The device is intended for use in general hospital settings and is operated by trained personnel under the direction of qualified medical staff. It is supplied as a reusable system requiring regular maintenance and is not designated as single-use or sterile. Storage and handling should follow manufacturer guidelines, and the device is not described as MRI-safe in the provided data.

Other devices from Ohmeda Medical

See all devices from Ohmeda Medical →

Questions

AI-generated answers, 26 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Ohio Infant Warmer System used for?

    The Ohio Infant Warmer System is designed to provide a controlled source of radiant heat for infants and pediatric patients. It supports both manual and servo modes of operation via a microprocessor-based control system, allowing clinicians to maintain appropriate thermal regulation in general hospital settings under trained staff supervision.

  • Is the Ohio Infant Warmer System a single-use or reusable device?

    The Ohio Infant Warmer System is supplied as a reusable system, meaning it is intended for repeated use after proper cleaning and maintenance. It is not designated as single-use, and regular maintenance is required to ensure safe and effective operation over time.

  • Is the Ohio Infant Warmer System sterile when supplied?

    The device is not described as sterile in the provided data. Since it is a reusable radiant warmer system that requires regular maintenance, it is not supplied in a sterile condition and should be prepared according to manufacturer guidelines before each use.

  • What are the storage and handling requirements for the Ohio Infant Warmer System?

    Storage and handling of the Ohio Infant Warmer System should follow the manufacturer’s guidelines to ensure device integrity and performance. Specific storage conditions are not detailed in the provided data, but users must adhere to the instructions provided by Ohmeda Medical for proper storage between uses.

  • What regulatory pathway was used for the Ohio Infant Warmer System?

    The Ohio Infant Warmer System received authorization through the FDA 510(k) process, as indicated by the FDA_510K designation in the device data. The decision was classified as 'Substantially Equivalent' (SESE) to a legally marketed predicate device, with a traditional clearance type granted on April 23, 1993.