Ohmeda Medical Giraffe Omnibed

by Ohmeda Medical

510(k) K020543

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK020543 Date9 May 2002ClearedStatusActive

Recalls

10 recalls · 4 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-2475-2026 FDA Class II initiated 1 May 2026
    • Ongoing Z-2477-2026 FDA Class II initiated 1 May 2026

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

    • Ongoing Z-0047-2025 FDA Class I initiated 3 Sep 2024

    GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

    • Ongoing Z-2101-2023 FDA Class II initiated 19 May 2023

    GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices

    • Terminated Z-2235-2020 FDA Class II initiated 28 Apr 2020

    Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.

    • Terminated Z-0538-2020 FDA Class I initiated 11 Oct 2019 on shared 510(k) K020543

    Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.

    • Terminated Z-1504-2017 FDA Class II initiated 18 Nov 2016 on shared 510(k) K020543

    Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

    • Terminated Z-1047-05 FDA Unclassified (pre-2012) initiated 25 Apr 2005

    Neonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.

    • Terminated Z-0288-04 FDA Unclassified (pre-2012) initiated 17 Oct 2003

    Heater vent door on neonatal incubator may malfunction.

    • Terminated Z-1064-03 FDA Unclassified (pre-2012) initiated 13 Jun 2003 on shared 510(k) K020543

    Firm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FMZ
Device Class
Class II
Regulation
21 CFR Part 880.5400
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Ohmeda Medical

See all devices from Ohmeda Medical →