Ohmeda Medical Giraffe Omnibed
510(k) K020543
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK020543 | Date9 May 2002 | StatusActive |
Recalls
-
- Ongoing Z-2475-2026
- Ongoing Z-2477-2026
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
-
- Ongoing Z-0047-2025
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
-
- Ongoing Z-2101-2023
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
-
- Terminated Z-2235-2020
Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
-
- Terminated Z-0538-2020
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
-
- Terminated Z-1504-2017
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
-
- Terminated Z-1047-05
Neonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.
-
- Terminated Z-0288-04
Heater vent door on neonatal incubator may malfunction.
-
- Terminated Z-1064-03
Firm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FMZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5400
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent