Ohmeda-Ohio Care Plus Incubator
510(k) K974349
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK974349 | Date12 Feb 1998 | StatusActive |
Recalls
-
- Ongoing Z-0957-2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
-
- Terminated Z-1510-2017
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FMZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5400
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent