Skip to content

Optima CT660

by GE Medical Systems, LLC 2 variants

510(k) K110227

Optima CT660 is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by GE Medical Systems, LLC, cleared under FDA 510(k) K110227 on 20 Jul 2011, active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK110227 Date20 Jul 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
25 Jan 2011
FDA decision
20 Jul 2011 (Substantially Equivalent)
FDA review time
176 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K110227 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

5 recalls · FDA

    • Terminated Z-2583-2018 FDA Class II initiated 24 May 2018

    Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

  • 2 recalls, same reason

    • Terminated Z-2871-2018 FDA Class II initiated 27 Apr 2018
    • Terminated Z-2873-2018 FDA Class II initiated 27 Apr 2018

    GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.

    • Terminated Z-2182-2017 FDA Class II initiated 19 May 2017

    It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.

    • Terminated Z-0023-2014 FDA Class II initiated 26 Mar 2013

    GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID

Timeline

6 dated events

  1. Cleared FDA 510(k) K110227
  2. Recall initiated FDA recalls Z-0023-2014
  3. Recall initiated FDA recalls Z-2182-2017
  4. Recall initiated FDA recalls Z-2871-2018
  5. Recall initiated FDA recalls Z-2873-2018
  6. Recall initiated FDA recalls Z-2583-2018

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAK
Device class
Class II
Regulation
21 CFR Part 892.1750
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K110227

1 device · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K110227
UDI-DISize and differences
NA Console Generation Upgrade
GE Hangwei Medical Systems Company Ltd.
ConsoleGenerationUpgrade / Console Generation Upgrade00195278882677 —On the market

Similar devices

Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).

Other devices from GE Medical Systems, LLC

See all devices from GE Medical Systems, LLC →