PASCAL Ace with Capture Clarity Valve Repair System 30000ACEMR
by Edwards Lifesciences LLC GUDID
Listed (no authorization)
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00690103221664 | Date24 Sep 2026 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2479-2023
Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
Specification
- Trade Name
- PASCAL Ace with Capture Clarity Valve Repair System
- Generic Name
- Heart valve clip
- Model / Ref / Catalog
- 30000ACEMR, 30000ACEMRA
- Primary DI / UDI-DI
- 00690103221664
- PMA Number
- FDA Product Code
- NKM
- GMDN Term
- Heart valve clip
- Device Class
- Class III
- Commercial Distribution
- In Commercial Distribution
- MRI Safety
- MR Conditional
Other devices from Edwards Lifesciences LLC
Properties
Other properties (7)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No