PASCAL Precision 20000GS
by Edwards Lifesciences LLC United States GUDID
Product code NKM
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00690103222388 | Date24 Sep 2026 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2479-2023
Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
Specification
- Trade Name
- PASCAL Precision
- Generic Name
- Heart valve clip
- Model / Ref / Catalog
- 20000GS, 20000GSLA
- Primary DI / UDI-DI
- 00690103222388
- PMA Number
- FDA Product Code
- NKM
- GMDN Term
- Heart valve clip
- Device Class
- Class III
- Commercial Distribution
- In Commercial Distribution
Other devices from Edwards Lifesciences LLC
Properties & raw data
Show properties, registry sources and raw records
Other properties (7)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
Sources (1)
- FDA UDI e561eb80-d631-48f9-8669-982c51444591