Pathfast CK-MB-II
Pathfast CK-MB-II is a medical device (FDA Class II; FDA product code JHX, Fluorometric Method, Cpk Or Isoenzymes) made by Phc Corporation, cleared under FDA 510(k) K081360 on 17 Aug 2009 and listed in FDA GUDID since 22 Sep 2016 (UDI-DI 04987595301339), active as of 21 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04987595301339 | Date22 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK081360 | Date17 Aug 2009 | StatusActive |
Official records
- AccessGUDID record 04987595301339
- FDA 510(k) database K081360
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K081360 decided 17 Aug 2009 Active
- FDA GUDID UDI-DI 04987595301339 listed 22 Sep 2016 Listed
Recalls
-
- Terminated Z-3185-2018
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Timeline
- Cleared FDA 510(k) K081360
- Listed FDA GUDID 04987595301339
- Recall initiated FDA recalls Z-3185-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Creatine kinase myocardial isoenzyme (CKMB) IVD, kit, chemiluminescent immunoassay
- Primary DI / UDI-DI
- 04987595301339
- 510(k) Number
- FDA product code
- JHX
- GMDN term
- Creatine kinase myocardial isoenzyme (CKMB) IVD, kit, chemiluminescent immunoassay
- Device class
- Class II
- Regulation number
- 862.1215
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Creatine kinase myocardial isoenzyme (CKMB) IVD, kit, chemiluminescent immunoassay.
- CK-MB AccuLite CLIA Test System - 192 Wells 2975-300 2975-300B by MONOBIND, INC.
- Dimension Vista® Flex® reagent cartridge MMB 10445097 K6420 by Siemens Healthcare Diagnostics Inc.
- Atellica IM CKMB 10995530 by Siemens Healthcare Diagnostics Inc.
- Monobind AccuLite CKMB CLIA Kit - 96 Wells 2975-300 2975-300A by MONOBIND, INC.