Skip to content

Plato 17-160-045

by Scientia Vascular, Inc.

Plato 17-160-045 (PL17-160-045) is a vascular microcatheter (FDA Class II) made by Scientia Vascular, Inc., cleared under FDA 510(k) K223560 on 21 Aug 2023 (1 earlier submission) and listed in FDA GUDID since 26 Jan 2023 (UDI-DI 00818075010462), active as of 23 Oct 2026. 2 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK223560 Date21 Aug 2023DecidedStatusActive
United StatesGUDID listing Number00818075010462 UDI-DI Date26 Jan 2023ListedStatusListed GUDID: in commercial distribution

Official records

From the FDA record

Device description
Microcatheter 45degree tip
MRI safety
MR Unsafe
Natural rubber latex
Labelled as not made with natural rubber latex
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K223560 decided 21 Aug 2023 Active· +1 more in the status table
  • FDA GUDID UDI-DI 00818075010462 listed 26 Jan 2023 Listed

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-1235-2025 FDA Class II initiated 16 Jan 2025

    Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

    • Ongoing Z-1249-2023 FDA Class II initiated 1 Feb 2023

    Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.

Timeline

5 dated events

  1. Cleared FDA 510(k) K210601
  2. Listed FDA GUDID 00818075010462
  3. Recall initiated FDA recalls Z-1249-2023
  4. Cleared FDA 510(k) K223560
  5. Recall initiated FDA recalls Z-1235-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Plato
Generic name
Vascular microcatheter
Model / Ref / Catalog
17-160-045, PL17-160-045
Primary DI / UDI-DI
00818075010462
510(k) Number
FDA product code
DQY, QJP
GMDN term
Vascular microcatheter
Device class
Class II
Regulation number
870.1250
GUDID distribution status
In Commercial Distribution
MRI safety
MR Unsafe
Size
Catheter Inner Diameter — 0.017 Inch
Catheter Working Length — 160 Centimeter
Pressure — 300 Pound per Square Inch
Outer Diameter — 0.57 Millimeter

Similar devices in this category

Other devices filed under Vascular microcatheter.

Other devices from Scientia Vascular, Inc.

  • Aristotle 01954

    Class II
  • Aristotle 01955

    Class II
  • Aristotle 01956

    Class II
  • Aristotle 01672

    Class II

See all devices from Scientia Vascular, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (10, all answered No)