Plato 17-160-090
Plato 17-160-090 (PL17-160-090) is a vascular microcatheter (FDA Class II) made by Scientia Vascular, Inc., cleared under FDA 510(k) K223560 on 21 Aug 2023 (1 earlier submission) and listed in FDA GUDID since 26 Jan 2023 (UDI-DI 00818075010479), active as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK223560 | Date21 Aug 2023 | StatusActive |
| United States | GUDID listing | Number00818075010479 | Date26 Jan 2023 | StatusListed |
Official records
- AccessGUDID record 00818075010479
- FDA 510(k) database K223560
- FDA 510(k) database K210601
From the FDA record
- Device description
- Microcatheter 90degree tip
- MRI safety
- MR Unsafe
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K223560 decided 21 Aug 2023 Active
- FDA GUDID UDI-DI 00818075010479 listed 26 Jan 2023 Listed
Recalls
-
- Ongoing Z-1235-2025
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
-
- Ongoing Z-1249-2023
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
Timeline
- Cleared FDA 510(k) K210601
- Listed FDA GUDID 00818075010479
- Recall initiated FDA recalls Z-1249-2023
- Cleared FDA 510(k) K223560
- Recall initiated FDA recalls Z-1235-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Plato
- Generic name
- Vascular microcatheter
- Model / Ref / Catalog
- 17-160-090, PL17-160-090
- Primary DI / UDI-DI
- 00818075010479
- FDA product code
- DQY, QJP
- GMDN term
- Vascular microcatheter
- Device class
- Class II
- Regulation number
- 870.1250
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Unsafe
- Size
- Catheter Inner Diameter — 0.017 InchCatheter Working Length — 160 CentimeterOuter Diameter — 0.57 MillimeterPressure — 300 Pound per Square Inch
Similar devices in this category
Other devices filed under Vascular microcatheter.
- Renegade™ HI-FLO™ FATHOM™ System M001184540 by Boston Scientific
- Merit Maestro® G 29MC29110SN by Merit Medical Systems, Inc.
- Merit Maestro® G 28MC21130SN by Merit Medical Systems, Inc.
- Direxion™ Transend™-14 System M001195870 by Boston Scientific
- PG PRO PGP28140STR by MICROVENTION INC.
- PG PRO PGP28165STR by MICROVENTION INC.